All In MedTech

Device Trial Radar · NCT07927231

Arduino-Based Real-Time Feedback for Manikin Endotracheal Intubation Training in Medical Students

Effects of an Arduino-Based Real-Time Feedback Device on Manikin-Based Endotracheal Intubation Training in Medical Students: A Cluster-Randomized Controlled Pretest-Posttest Study

Last checked: Nov 19, 2026Latest source record: Nov 10, 2026 (this record)Source: ClinicalTrials.govMethods & limitations

Completed

Status

Completed

Enrollment

80

Actual

Start

Aug 16, 2022

Primary completion

Aug 25, 2023

Study completion Aug 26, 2023

Sponsor

Lead sponsor
Tri-Service General Hospital (TSGH)Academic, hospital or other
Sponsor type
Academic, hospital or other
Collaborators
National Defense Medical University
First posted
Nov 10, 2026

More device trials from this sponsor

Design

Study type
Interventional
Phase
Not applicable (typical for device studies)
Allocation
Randomized
Model
Parallel
Masking
None
Primary purpose
Other
FDA-regulated device
No
Who can join
All sexes, from 18 years
Healthy volunteers
Yes

Summary

As written by the sponsor on the registry.

This study evaluated whether an Arduino-based device that provides real-time pressure and time feedback during manikin practice improves endotracheal intubation training for sixth-year medical students. Small teaching groups of three to four students were allocated by lot to practice either with a feedback-equipped manikin or with a standard manikin, within the same structured airway-management course. Upper-incisor loading during laryngoscopy, learner-reported autonomy, competence, intubation self-efficacy, and course satisfaction were measured before and after training, and intubation success was recorded during subsequent operating-room practice.

Conditions

  • Endotracheal Intubation
  • Medical Education
  • Simulation Training

Interventions

  • Arduino-based real-time feedback laryngoscopeDevice

    Standard laryngoscope blade fitted with a pressure-sensing element at the tooth end, connected to an Arduino-based display that shows a unitless upper-incisor pressure score and elapsed intubation time after each manikin intubation attempt, with an audible alarm when sensed pressure exceeds a preset threshold. Used during coached manikin practice within the structured airway-management course.

  • Structured airway-management courseBehavioral

    Course-opening concept map of airway management, direct teaching and demonstration of oral and nasal airways, bag-mask ventilation, and endotracheal intubation, followed by coached practice on an airway manikin with instructor feedback, reflection, and reattempts. Delivered identically to both arms.

Primary outcomes

What the sponsor says the trial is designed to measure.

  1. Upper-incisor pressure scoreTime frame: Before training (pretest) and immediately after training (posttest), within the same teaching session

Locations

1 site in 1 country. Site status is shown where the registry gives one.

  • Tri-Service General Hospital

    Taipei, Taiwan

Contact details for sites and study staff are not shown here. To ask about taking part, use the contact information on the registry record.

For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.