Noninvasive Blood Pressure Validation Study

Last checked: Nov 21, 2026Latest source record: Aug 19, 2026 (this record)Source: ClinicalTrials.govMethods & limitations

Completed

Status

Completed

Enrollment

196

Actual

Start

Jan 15, 2026

Primary completion

Jun 15, 2026

Study completion Jun 15, 2026

Sponsor

Lead sponsor
Perin Health DevicesNot on our public-company list
Sponsor type
Industry
First posted
Aug 19, 2026

More device trials from this sponsor

Design

Study type
Observational
FDA-regulated device
Yes
Who can join
All sexes, from 22 years, up to 90 years
Healthy volunteers
Yes

Summary

As written by the sponsor on the registry.

The purpose of this study is to test how accurately the Perin Health Patch, a small wearable sensor worn on the chest, measures blood pressure without using an arm cuff. Adults 22 to 90 years of age wore the patch while trained observers took blood pressure readings at the same time using a standard arm cuff and stethoscope. The patch readings were then compared to the standard readings to see how closely they match, following recognized international testing standards for blood pressure devices (ISO 81060-2 and IEEE 1708). Participants visited one of three study sites up to three times over about one week. At the final visit, participants also did light resistance band exercise to briefly raise their blood pressure and relaxation activities to lower it, so the patch could be tested while blood pressure was changing.

Conditions

  • Blood Pressure Monitoring
  • Blood Pressure

Interventions

  • Chest Worn Blood Pressure MonitorDevice

    The Perin Health Patch (PHP) is an investigational, battery powered, wearable physiological monitoring device with an adhesive patch and sensor module, applied to the upper left chest. Cuffless noninvasive blood pressure measurements from the device were compared to a reference sphygmomanometer (dual observer auscultation, same arm simultaneous method).

Primary outcomes

What the sponsor says the trial is designed to measure.

  1. Mean Error of Paired Blood Pressure Determinations (ISO 81060-2 Criterion 1)Time frame: Day 1 through Day 7 (visit window up to Day 9 per participant)

3 more outcomes with a free account

Locations

1 site in 1 country. Site status is shown where the registry gives one.

  • Element Boulder

    Louisville, Colorado

Contact details for sites and study staff are not shown here. To ask about taking part, use the contact information on the registry record.

For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.