All In MedTech

Device Trial Radar · NCT07768215

Miniature In-line Respirator

First-In-Human Trial of Miniature In-Line, Pressure-Cycled, Emergency Respirator Device in Human Organ Donors

Last checked: Nov 19, 2026Latest source record: Nov 19, 2026 (this record)Source: ClinicalTrials.govMethods & limitations

Not yet recruiting

Status

Not yet recruiting

Enrollment

5

Planned (sponsor's estimate)

Start

Nov 24, 2026

Primary completion

Dec 1, 2028

Study completion Dec 1, 2028

Sponsor

Sponsor type
NIH
First posted
Aug 17, 2026

More device trials from this sponsor

Design

Study type
Interventional
Phase
Not applicable (typical for device studies)
Allocation
Not applicable (single group)
Model
Single group
Masking
None
Primary purpose
Device feasibility
FDA-regulated device
No
Who can join
All sexes, from 1 year, up to 100 years
Healthy volunteers
No

Summary

As written by the sponsor on the registry.

This first-in-human pilot study will test an investigational miniature breathing device called the Hope inVent. The device is a small, single-use, in-line respirator that connects to a breathing tube and uses compressed oxygen to help move air in and out of the lungs. It has no moving parts and is designed to provide short-term breathing support in situations where standard ventilators may not be available, such as emergencies, disasters, battlefield settings, transport, or ventilator shortages. The study will enroll up to 5 brain-dead organ donors who are already receiving mechanical ventilation at the Gift of Life Donor Care Center at the Hospital of the University of Pennsylvania. Donors will only be included after authorization for organ donation and research has been obtained through standard procedures. Donors who may donate lungs or who have significant lung disease will not be included. During the study, investigators will briefly replace the standard ventilator with the Hope inVent device for up to 15 minutes. The standard ventilator will remain immediately available, and the donor will be continuously monitored by the clinical and research teams. Investigators will measure whether the device can maintain acceptable breathing measures, including oxygen level and carbon dioxide level. They will also record breathing pressures and tidal volume, which is the amount of air delivered with each breath. If the device does not perform as expected or if any safety concern occurs, the donor will be returned immediately to the standard ventilator. There is no direct benefit to the organ donor. Information from this study may help determine whether this type of simple, low-cost breathing device could be useful for future patients who need emergency or short-term ventilator support when standard ventilators are not available.

Conditions

  • Brain Death
  • Respiration, Artificial

Interventions

  • Hope inVentDevice

    The Hope inVent is an investigational, single-use, miniature in-line respirator designed to provide short-term ventilatory support for intubated individuals. The device is a 3D-printed, pressure-cycled, gas-powered respirator with no moving parts. It connects to the endotracheal tube breathing circuit and uses compressed oxygen as its power source. In this study, only the Green Hope inVent model will be used. The device cycles between fixed pressure settings: peak inspiratory pressure of 18 cm H2O and positive end-expiratory pressure of 6 cm H2O. Enrolled brain-dead research donors will be placed on the device once for up to 15 minutes, with continuous respiratory and physiologic monitoring. The donor will then be returned to the standard mechanical ventilator.

Primary outcomes

What the sponsor says the trial is designed to measure.

  1. Proportion of donors with successful ventilation and oxygenation during Hope inVent testing.Time frame: During device use, from connection of the Hope inVent device to removal, up to 15 minutes.

Locations

1 site in 1 country. Site status is shown where the registry gives one.

  • Hospital of the University of Pennsylvania

    Philadelphia, Pennsylvania

Contact details for sites and study staff are not shown here. To ask about taking part, use the contact information on the registry record.

For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.