A Three-phase Study That Will Compare the ECG Data Recorded Using the Test Device With the Data Recorded by a Reference Device, Evaluate the ECG Signal Quality of the Test Device Over a 10-day Simulated Use Period, and Assess the Effects of Variable ECG Electrode Placement for the Test Device

Clinical Assessment of ECG Signal Quality With the ZywieZ3 Sensor and Adhesive Versus a Traditional Device Over a 1 Hour Period, Monitoring Skin Adhesion / ECG Performance of the ZywieZ3 Sensor and Adhesive Over 10 Days, and of the Effects of Variable ECG Electrode Placement With the ZywieZ3 Sensor and Adhesive

Last checked: Nov 20, 2026Latest source record: Oct 1, 2025 (this record)Source: ClinicalTrials.govMethods & limitations

Completed

Status

Completed

Enrollment

45

Actual

Start

Jan 15, 2025

Primary completion

Feb 14, 2025

Study completion Feb 14, 2025

Sponsor

Lead sponsor
Zywie, Inc.Not on our public-company list
Sponsor type
Industry
First posted
Oct 1, 2025

More device trials from this sponsor

Design

Study type
Interventional
Phase
Not applicable (typical for device studies)
Allocation
Non randomized
Model
Single group
Masking
None
Primary purpose
Diagnostic
FDA-regulated device
Yes
Who can join
All sexes, from 18 years, up to 80 years
Healthy volunteers
Yes

Summary

As written by the sponsor on the registry.

This clinical study will be performed in three phases. The purpose of Phase 1 of this study is to compare the ECG data recorded using the ZywieZ3 Sensor and ZywieZ3 Adhesive ("Test Device") to the data recorded by a traditional 3-lead ECG device that uses wet Ag-AgCl adhesives in a standard lead location ("Reference Device"). The ECG signals will be recorded simultaneously with both ECG devices from the subjects in conditions that represent the devices' intended use for 1 hour. This comparison will determine whether the Test Device recordings are similar to the "gold standard" Reference Device. The testing methodology for this portion of the study is based on guidance provided by the FDA to the Sponsor. The purpose of Phase 2 of this study is to evaluate the ECG signal quality of the Test Device over a 10-day simulated use period by human subjects. This duration of 10 days represents the expected maximum period of usage for the Test Device by the Sponsor. The skin adhesion properties of the Test Device will be evaluated, including a determination of whether any skin reactions, discoloration of the Test Device adhesive (including the snaps and hydrogel), or Test Device adhesive degradation leading to particulate formation occurs over a 10-day simulated use period by human subjects. During this testing, a human factor critical task analysis will be conducted via a collection of observational data regarding success in subject completion of Sponsor-identified critical tasks, subject completion of knowledge questionnaires relating to aspects of Sponsor-identified critical tasks, and subject completion of general subjective questionnaires regarding any difficulty they experienced with use of the Test Device. Testing methodology for this portion of the study is based upon the "Adhesive Performance (Duration of Use)" \[Section 5.4\] procedures from ANSI/AAMI Standard EC12:2000/(R)2020, Disposable ECG Electrodes and the 2016 FDA Guidance for Industry and Food and Drug Administration Staff titled "Applying Human Factors and Usability Engineering to Medical Devices." The purpose of Phase 3 of this clinical study is to assess the effects of variable ECG electrode placement for the ZywieZ3 Sensor and ZywieZ3 Adhesive ("Test Device"). This testing is intended to determine whether ECG electrode placement variability produces unacceptable / non-clinically relevant ECG data and what level of ECG electrode placement variability is acceptable.

Conditions

  • Healthy Subjects

Interventions

  • Phase 1 Test Device Sensor and Adhesive and Reference DeviceDevice

    Test Device Sensor and Adhesive and Reference Device ECG Data Collection for 1 hour

  • Phase 2 Test Device ECG Signal quality over a 10-day simulated use periodDevice

    Evaluation of the ECG-signal quality of the Test Device sensor and adhesive applied to a single site on a subject's chest to collect ECG readings for a 10-day period.

  • Test Device Sensor and Adhesive Variable PlacementDevice

    Comparison of nine incorrect locations of Test Device Sensor and Adhesive to the ideal location

Primary outcomes

What the sponsor says the trial is designed to measure.

  1. Phase 1: Test Device Sensor and Adhesive Equivalence to a Reference DeviceTime frame: From enrollment to end of the 1-hour ECG reading period

2 more outcomes with a free account

Locations

1 site in 1 country. Site status is shown where the registry gives one.

  • Bioscience of Bozeman, LLC

    Bozeman, Montana

Contact details for sites and study staff are not shown here. To ask about taking part, use the contact information on the registry record.

For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.