Device Trial Radar · NCT06503354
Gait Metric Validation of the Cionic Neural Sleeve
Last checked: Nov 19, 2026Latest source record: Dec 18, 2025 (this record)Source: ClinicalTrials.govMethods & limitations
Status
Completed
Enrollment
29
Actual
Start
Jul 8, 2024
Primary completion
Dec 1, 2025
Study completion Dec 11, 2025
Sponsor
- Lead sponsor
- Cionic, Inc.Not on our public-company list
- Sponsor type
- Industry
- First posted
- Jul 16, 2024
Design
- Study type
- Observational
- FDA-regulated device
- Yes
- Who can join
- All sexes, from 22 years, up to 64 years
- Healthy volunteers
- Yes
Summary
As written by the sponsor on the registry.
The purpose of this study is to determine the validity of the Neural Sleeve gait metrics compared to gait parameters assessed with a Protokinetics Zeno pressure sensitive walkway. The spatiotemporal gait parameters obtained from the Zeno walkway have been rigorously validated and will represent the "gold standard" ground truth measurements in the proposed protocol. Confirming the validity of the Neural Sleeve metrics offers benefit to individuals who utilize the device by providing confidence that calculated gait reports are accurate and can be used to inform usage of the device and associated gait changes over time. Findings of inaccuracy would still offer benefit in the form of knowing that the Neural Sleeve metrics need to be used with caution and/or updated to provide correct outcomes to users.
Conditions
- Validation Study
Interventions
Neural Sleeve and Zeno walkwayDevice
A standard manual stopwatch will be used to record time. The Zeno pressure gait mat will be laid flat on the ground and used for all walking conditions/speed. Participants will perform two 5-minute walking bouts for each of the four walking conditions while wearing the Neural Sleeve without electrical stimulation. For normal walking speed, participants walked at their normal, comfortable pace for 5 minutes. For fast walking speed, participants walked to cover as much distance as possible for 5 minutes. For intermittent-slow speed, participants were instructed to walk as if browsing in a museum. For slow speed, participants walked at a slow, steady pace. The order of conditions was randomized. They will walk back and forth along a 12-meter course with the gait mat in the center of the course. Participants will not receive any intervention or treatment. Simultaneously, the Zeno walkway will be operated according to manufacturer's instructions.
Primary outcomes
What the sponsor says the trial is designed to measure.
- Gait speedTime frame: A walking session of up to 5 minutes
- Stride lengthTime frame: A walking session of up to 5 minutes
- CadenceTime frame: A walking session of up to 5 minutes
Locations
1 site in 1 country. Site status is shown where the registry gives one.
Cleveland State University, Department of Health Sciences and Human Performance
Cleveland, Ohio
Contact details for sites and study staff are not shown here. To ask about taking part, use the contact information on the registry record.
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.