Device Trial Radar · NCT06422325
Two Way Crossover Closed Loop Study Insulin vs Insulin and Pramlintide
Last checked: Nov 20, 2026Latest source record: Jun 9, 2026 (this record)Source: ClinicalTrials.govMethods & limitations
Status
Completed
Enrollment
33
Actual
Start
Jul 12, 2024
Primary completion
Jan 4, 2025
Study completion Jan 28, 2025
Sponsor
- Lead sponsor
- Oregon Health and Science UniversityAcademic, hospital or other
- Sponsor type
- Academic, hospital or other
- Collaborators
- National Institute of Diabetes and Digestive and Kidney Diseases (NIDDK)
- First posted
- May 20, 2024
Design
- Study type
- Interventional
- Phase
- Not applicable (typical for device studies)
- Allocation
- Randomized
- Model
- Crossover
- Masking
- None
- Primary purpose
- Treatment
- FDA-regulated device
- Yes
- Who can join
- All sexes, from 18 years, up to 70 years
- Healthy volunteers
- No
Summary
As written by the sponsor on the registry.
The purpose of this study is to test how well a new investigational closed loop system manages your blood sugar with the ability to deliver insulin and pramlintide. Pramlintide is a drug that is used with mealtime insulin to control blood sugar in people who have diabetes. It works by slowing down the movement of food through the stomach which prevents blood sugar from rising too high after a meal. The closed loop system will receive glucose values from the Dexcom G6 continuous glucose monitoring (CGM) and automatically send commands to one Omnipod for insulin and one Omnipod for pramlintide delivery.
Conditions
- Type 1 Diabetes
Interventions
MPC closed-loop system in insulin only modeDevice
The Model Predictive Control (MPC) insulin infusion algorithm contains a model within the controller that takes as an input the aerobic metabolic expenditure in addition to the CGM and meal in puts. The algorithm uses heart rate and accelerometer data collected on the patient's body to calculate metabolic expenditure (METs). The METs then acts on the model for the insulin dynamics, whereby more energy expenditure and longer duration exercise can lead to a more substantial effect of insulin on the CGM. The MPC also has missed meal insulin bolus detection where the system will calculate the amount of insulin that was missed for a meal. The missed meal boluses can be delivered automatically without any input from the user. This feature can also be disabled. The device in this mode will administer insulin continuously for managing blood sugar.
MPC closed-loop system in insulin/pramlintide modeDevice
The Model Predictive Control (MPC) insulin infusion algorithm contains a model within the controller that takes as an input the aerobic metabolic expenditure in addition to the CGM and meal in puts. The algorithm uses heart rate and accelerometer data collected on the patient's body to calculate metabolic expenditure (METs). The METs then acts on the model for the insulin dynamics, whereby more energy expenditure and longer duration exercise can lead to a more substantial effect of insulin on the CGM. The MPC also has missed meal insulin bolus detection where the system will calculate the amount of insulin that was missed for a meal. The missed meal boluses can be delivered automatically without any input from the user. This feature can also be disabled. The device in this mode will administer both insulin and pramlintide continuously for managing blood sugar. The system will deliver pramlintide in a fixed ratio to insulin at 6 mcg of pramlintide delivered for every 1 unit of insulin.
Primary outcomes
What the sponsor says the trial is designed to measure.
- Incremental Area Under the Curve of Postprandial Glucose Following the First MealTime frame: 6 hours following first meal
- Percent of Time With Sensed Glucose Between 70 - 180 mg/dl Following First MealTime frame: 6 hours following first meal
Locations
1 site in 1 country. Site status is shown where the registry gives one.
Oregon Health and Science University
Portland, Oregon
Contact details for sites and study staff are not shown here. To ask about taking part, use the contact information on the registry record.
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.