All In MedTech

Device Trial Radar · NCT06203795

Dialysis Performance of the FX CorAL Membrane

Performance of Toxin Removal and Clotting Kinetics in Dialyzers

Last checked: Nov 20, 2026Latest source record: Mar 10, 2025 (this record)Source: ClinicalTrials.govMethods & limitations

Completed

Status

Completed

Enrollment

10

Actual

Start

Oct 2, 2023

Primary completion

Oct 26, 2023

Study completion Dec 1, 2024

Sponsor

Lead sponsor
University Hospital, GhentAcademic, hospital or other
Sponsor type
Academic, hospital or other
First posted
Jan 12, 2024

More device trials from this sponsor

Design

Study type
Interventional
Phase
Not applicable (typical for device studies)
Allocation
Randomized
Model
Crossover
Masking
None
Primary purpose
Treatment
FDA-regulated device
No
Who can join
All sexes, from 18 years
Healthy volunteers
No

Summary

As written by the sponsor on the registry.

Anticoagulation remains an important issue in the setting of hemodialysis, and up till now there are some major points on which further research is needed. First, it is important to have a portfolio of the performance of different commercially available dialyzers with respect to fiber clotting. Second, to better estimate the impact of clotting on the overall dialysis performance, clotting kinetics during dialysis should be understood. The aim of the present project is therefore to quantify the performance of the FX CorAL dialyzer (Fresenius Medical Care, Germany) in settings with reduced anticoagulation, and compare different performance outcomes (percent open fibers, solute removal rates) to those of other commercially available dialyzers. The different outcomes are related to the dialyzer extraction ratio and reduction ratio of small and middle molecules and albumin, the visual scoring of the dialyzer post dialysis, and the anticoagulation properties as assessed by fiber blocking in the dialyzer. Performance parameters at different time steps will allow to better understand clotting kinetics during dialysis.

Conditions

  • Hemodialysis Complication
  • Anticoagulants

Interventions

  • Anticoagulation dosingOther

    Patients are dialyzed with only one quarter of their regular anticoagulation dose during the sessions on Tuesday and Thursday

  • Blood samplingOther

    Blood is sampled at start and at 10, 30, 60, 180, and 240 min after dialysis start for the 4 hours dialysis session, and at start and 10 and 180 min after dialysis start for the 3 hours dialysis session.

  • Duration of the dialysis sessionOther

    Patients have a washout with their regular dialysis duration on Saturday, a study session of 4 hours dialysis on Tuesday, and a study session of 3 hours dialysis on Thursday

  • Choice of the hemodialyzerDevice

    Patients are dialyzed with one of the four dialyzers during the entire week, starting at Saturday (i.a. Saturday, Tuesday and Thursday). Block randomization is performed per two weeks. In week 1 and 2, patients are dialyzed with the FX CorAL 800 and the FX 800 Cordiax. In week 3 and 4 of the study, patients are dialyzed with the Solacea 19H and the Xevonta Hi20.

Primary outcomes

What the sponsor says the trial is designed to measure.

  1. Fiber patency (i.e. number of open dialyzer fibers) in FX CorAL versus the competing dialyzersTime frame: 4 months

1 more outcomes with a free account

Locations

1 site in 1 country. Site status is shown where the registry gives one.

  • Ghent University Hospital

    Ghent, Belgium

Contact details for sites and study staff are not shown here. To ask about taking part, use the contact information on the registry record.

For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.