A Clinical Trial to Verify the Accuracy of Seers Technology's Pulse Oximeter (mobiCARE +Pulse, MP100W) in Healthy Volunteers
Last checked: Nov 20, 2026Latest source record: Jun 8, 2023 (this record)Source: ClinicalTrials.govMethods & limitations
Status
Completed
Enrollment
12
Actual
Start
Apr 29, 2023
Primary completion
May 14, 2023
Study completion May 15, 2023
Sponsor
- Lead sponsor
- Seers Technology Co., Ltd.Not on our public-company list
- Sponsor type
- Industry
- First posted
- Jun 8, 2023
Design
- Study type
- Interventional
- Phase
- Not applicable (typical for device studies)
- Allocation
- Not applicable (single group)
- Model
- Single group
- Masking
- None
- Primary purpose
- Other
- FDA-regulated device
- No
- Who can join
- All sexes, from 20 years, up to 50 years
- Healthy volunteers
- Yes
Summary
As written by the sponsor on the registry.
It is important to evaluate the accuracy of the pulse oximeter, which is commonly used in clinical practice. The US FDA only allows the use of devices that meet the ISO 80601-2-61:2017 standard in the United States. According to the ISO 80601-2-61:2017 standard, the accuracy of the oxygen saturation (SpO2) of the pulse oximeter should have an error range of less than 4.0% compared to the actual arterial blood oxygen saturation (SaO2) in the range of 70-100%. To prove compliance with this requirement, it is necessary to conduct a clinical trial that induces hypoxemia in healthy adults and compares and evaluates the arterial blood oxygen saturation values of the pulse oximeter's SpO2 and the arterial blood oxygen saturation values of the carbon monoxide-oxygen meter (CO-oximeter).
Conditions
- Oxygen Saturation
Interventions
mobiCARE +Pulse (MP100W)Device
Volunteers were monitored using end-tidal carbon dioxide partial pressure and fraction of inspired oxygen. Each volunteer was placed in a semi-Fowler's position and connected to a breathing circuit to administer the nitrogen-air-carbon dioxide mixtures. A nose clip was applied to prevent breathing of room air. For frequent blood sampling, an arterial cannula was placed in the radial artery of each volunteer. Pulse oximeter probes were simultaneously attached to each volunteer's fingers. Each volunteer was exposed to various levels of induced hypoxia from 70\~100% of SaO2. Each plateau of oxygen saturation was maintained for at least 30 s until stabilization, after which 1 ml of arterial blood was drawn into a heparinized syringe. The study period consisted of two rounds of hypoxia, and the volunteers were maintained on room air between each round. SaO2 measurements using a CO-oximeter were used as a reference for the SpO2 accuracy.
Primary outcomes
What the sponsor says the trial is designed to measure.
- Accuracy of SpO2Time frame: 2 day
Locations
1 site in 1 country. Site status is shown where the registry gives one.
Seers Technology Co., Ltd.
Pyeongtaek-si, Gyeonggi-do, South Korea
Contact details for sites and study staff are not shown here. To ask about taking part, use the contact information on the registry record.
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.