Investigation of a Transcatheter Tricuspid Valved Stent Graft in Patients With Carcinoid Heart Disease

Single Centre, Prospective, Single Arm Open Label Series of a Transcatheter Tricuspid Valved Stent Graft System for the Reduction of Tricuspid Regurgitation and Improved Patient Reported Outcomes in Patients With Carcinoid Heart Disease

Last checked: Nov 19, 2026Latest source record: Nov 19, 2026 (this record)Source: ClinicalTrials.govMethods & limitations

Active, not recruiting

Status

Active, not recruiting

Enrollment

15

Planned (sponsor's estimate)

Start

Apr 6, 2022

Primary completion

May 21, 2028

Study completion May 21, 2028

Sponsor

Lead sponsor
Queen Mary University of LondonAcademic, hospital or other
Sponsor type
Academic, hospital or other
Collaborators
Royal Free Hospital NHS Foundation Trust
First posted
Oct 1, 2021

More device trials from this sponsor

Design

Study type
Interventional
Phase
Not applicable (typical for device studies)
Allocation
Not applicable (single group)
Model
Single group
Masking
None
Primary purpose
Treatment
FDA-regulated device
No
Who can join
All sexes, from 18 years
Healthy volunteers
No

Summary

As written by the sponsor on the registry.

The purpose of this investigation is to see if the TRICENTO Valved Stent Graft implant reduces tricuspid regurgitation (TR) and improves the symptoms and quality of life in 15 participants with carcinoid heart disease, and who are not able to have a new valve via a surgical procedure.

Conditions

  • Tricuspid Regurgitation
  • Tricuspid Valve Disease
  • Carcinoid Syndrome
  • Carcinoid Heart Disease

Interventions

  • Transcatheter Tricuspid Valved Stent GraftDevice

    All participants meeting the inclusion and none of the exclusion criteria will be enrolled. A custom-made Transcatheter Tricuspid Valved Stent Graft based on CT to define anatomic dimensions will be made. The system consists of a custom-made valved stent graft delivery system, a custom-made valved stent graft bioprosthesis and a loading system. The bioprosthesis is self-expanding and consists of a stent graft spanning from the inferior vena cava to the superior vena cava and a lateral bicuspid valve element made of thin porcine pericardial leaflets requiring a low closing pressure, and Nitinol support structures. The participants will undergo a minimally invasive, catheter-based procedure to implant the bioprosthesis which is delivered through a 24 French delivery system via transfemoral access. The device can be repositioned and is retrievable.

Primary outcomes

What the sponsor says the trial is designed to measure.

  1. The number of patients with successful implantation of the TRICENTO bioprosthesisTime frame: measured pre intervention and immediately after the intervention

Locations

1 site in 1 country. Site status is shown where the registry gives one.

  • Barts Health NHS Trust

    London, England, United Kingdom

Contact details for sites and study staff are not shown here. To ask about taking part, use the contact information on the registry record.

For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.