Device Trial Radar · NCT04684862
Infusion System for Hepatic Cancer
Last checked: Nov 20, 2026Latest source record: Dec 15, 2025 (this record)Source: ClinicalTrials.govMethods & limitations
Status
Active, not recruiting
Enrollment
100
Planned (sponsor's estimate)
Start
Feb 8, 2021
Primary completion
Jul 2029
Study completion Jul 2029
Sponsor
- Lead sponsor
- Ronald DeMatteo, M.D.Academic, hospital or other
- Sponsor type
- Academic, hospital or other
- First posted
- Dec 28, 2020
Design
- Study type
- Interventional
- Phase
- Not applicable (typical for device studies)
- Allocation
- Not applicable (single group)
- Model
- Single group
- Masking
- None
- Primary purpose
- Treatment
- FDA-regulated device
- Yes
- Who can join
- All sexes, from 18 years
- Healthy volunteers
- No
Summary
As written by the sponsor on the registry.
This is a single-site, open-label continued access study/treatment protocol under a treatment IDE. In addition to treating patients, the primary objective of this study is to assess the safety of using the Medtronic SynchroMed II programmable pump combined with the Intera tapered catheter for hepatic artery infusion (HAI) of a standard chemotherapy (FUDR) drug for adults with a clinical or biopsy-proven diagnosis of colorectal cancer metastatic to the liver or intrahepatic cholangiocarcinoma. After successful implantation, the combined pump and catheter system will be evaluated using a nuclear scan in the postoperative period, which is standard procedure to confirm that the pump is functioning prior to HAI of FUDR. Monitoring for safety will include a record of residual pump volume when it is emptied (every 2-12 weeks depending on whether the pump is being used for chemotherapy infusion) to determine if the pump is still working and surveillance of routine cross-sectional imaging (usually every 2-6 months) for any sign of a pump or catheter problem. Patients will be monitored for the safety of the pump/catheter combination for up to 5 years or pump removal/study withdrawal.
Conditions
- Colorectal Cancer Metastatic to Liver
- Intrahepatic Cholangiocarcinoma
Interventions
Medtronic SynchroMed II programmable pump connected to an Intera tapered catheter (Combined Infusion System)Device
This infusion system is composed of a Medtronic SynchroMed II programmable pump connected to a Intera tapered catheter. Specifically, the Medtronic pump comes with a small connector that ends in a metal flange. During the operation, under sterile conditions, the distal 1 cm of the flange is severed using a scissor. Then, using a Intera connector pin, the end of the Medtronic catheter is connected to the Intera catheter . Two 2-0 silk ties secure the catheters to the pin. The pump and catheter are then inserted into the patient into the abdominal wall or into the chest wall. The catheter is inserted into an artery within the abdomen (usually the gastroduodenal artery but it depends on the exact arterial branching in the individual patient) that feeds the liver.
Primary outcomes
What the sponsor says the trial is designed to measure.
- Number of Targeted Adverse Events (TAEs)Time frame: From surgery up to 60 months post-surgery or study withdrawal
Locations
1 site in 1 country. Site status is shown where the registry gives one.
Hospital of the University of Pennsylvania
Philadelphia, Pennsylvania
Contact details for sites and study staff are not shown here. To ask about taking part, use the contact information on the registry record.
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.