All In MedTech

Device Trial Radar · NCT04482556

Comparison of the Q-NRG+ Indirect Calorimetry Device Versus the V(Max) Encore Device in Mechanically Ventilated Children

Last checked: Nov 20, 2026Latest source record: Jan 3, 2024 (this record)Source: ClinicalTrials.govMethods & limitations

Status unknown

Status

Status unknown

Enrollment

40

Planned (sponsor's estimate)

Start

Jan 1, 2020

Primary completion

Dec 1, 2024

Study completion Dec 31, 2024

Sponsor

Lead sponsor
Boston Children's HospitalAcademic, hospital or other
Sponsor type
Academic, hospital or other
First posted
Jul 22, 2020

More device trials from this sponsor

Design

Study type
Interventional
Phase
Not applicable (typical for device studies)
Allocation
Non randomized
Model
Single group
Masking
None
Primary purpose
Diagnostic
FDA-regulated device
Yes
Who can join
All sexes, from 10 years, up to 18 years
Healthy volunteers
No

Summary

As written by the sponsor on the registry.

The overarching aim of this proposal is to examine the feasibility of the Q-NRG+ indirect calorimetry device and its agreement with (Vmax) Encore indirect calorimetry device in mechanically ventilated children. The overall hypothesis of this study is that the Q-NRG+ will provide minute-to-minute oxygen consumption (VO2) and carbon dioxide production (CO2) measurements that are in agreement with those obtained by the standard indirect calorimetry device currently used at our institution (Vmax Encore).

Conditions

  • Calorimetry, Indirect
  • Oxygen Consumption
  • Metabolism
  • Mechanical Ventilation

Interventions

  • Q-NRG+ Indirect Calorimetry DeviceDevice

    The two indirect calorimetry devices will be tested in an alternating fashion for each enrolled patient. Each enrolled patient subsequently will alternate in which device is tested first.

  • V(max) Encore Indirect Calorimetry DeviceDevice

    The two indirect calorimetry devices will be tested in an alternating fashion for each enrolled patient. Each enrolled patient subsequently will alternate in which device is tested first.

Primary outcomes

What the sponsor says the trial is designed to measure.

  1. Mean bias and limits of agreement between measured oxygen consumption (VO2) and carbon dioxide elimination (VCO2) in mechanically ventilated children assessed using the Q-NRG+ indirect calorimetry device and the V(max) Encore indirect calorimetry device.Time frame: 1 hour/participant
  2. Calculated total testing time, including calibration, required to complete a steady state indirect calorimetry measurement using the Q-NRG+ device and the V(max) Encore.Time frame: 1 hour/participant

Locations

1 site in 1 country. Site status is shown where the registry gives one.

  • Boston Children's Hospital

    Boston, Massachusetts · Recruiting

Contact details for sites and study staff are not shown here. To ask about taking part, use the contact information on the registry record.

For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.