All In MedTech

Device Trial Radar · NCT04360369

Comparison of Intraocular Pressure Measurements Between Reichert Tono-Vera Tonometer and Goldmann Tonometry

Last checked: Nov 19, 2026Latest source record: Jan 3, 2022 (this record)Source: ClinicalTrials.govMethods & limitations

Completed

Status

Completed

Enrollment

61

Actual

Start

Oct 19, 2020

Primary completion

Apr 28, 2021

Study completion Apr 28, 2021

Sponsor

Lead sponsor
Reichert, Inc.Not on our public-company list
Sponsor type
Industry
First posted
Apr 24, 2020

More device trials from this sponsor

Design

Study type
Interventional
Phase
Not applicable (typical for device studies)
Allocation
Non randomized
Model
Sequential
Masking
None
Primary purpose
Diagnostic
FDA-regulated device
Yes
Who can join
All sexes, from 18 years, up to 90 years
Healthy volunteers
Yes

Summary

As written by the sponsor on the registry.

Determine if the Tono-Vera Tonometer accurately measures intraocular pressure (IOP). The hypothesis of this test is to confirm the Tono-Vera Tonometer is equivalent to the Goldmann Applanation Tonometer (within +/- 5.0 mmHg).

Conditions

  • Intraocular Pressure
  • Glaucoma

Interventions

  • IOP with Goldmann Applanation TonometerDevice

    Measurement of intraocular pressure (IOP) with Goldmann Applanation Tonometer. Measurement will be used to categorize each subject as having Low IOP (7 to 16 mmHg), Medium IOP (\>16 to \<23 mmHg), or High IOP (\>23 mmHg).

  • IOP with comparator ORA G3 and ic100 tonometersDevice

    Measurement of IOP with Ocular Response Analyzer G3 and ic100 tonometers.

  • IOP with Tono-Vera TonometerDevice

    Measurement of IOP with Tono-Vera Tonometer

Primary outcomes

What the sponsor says the trial is designed to measure.

  1. Demonstrate compliance with ANSI Z80.10-2014-Ophthalmics-Ophthalmic Instruments-Tonometers.Time frame: Through study completion, approximately 4 months.

Locations

1 site in 1 country. Site status is shown where the registry gives one.

  • Department of Ophthalmology, Duke University

    Durham, North Carolina

Contact details for sites and study staff are not shown here. To ask about taking part, use the contact information on the registry record.

For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.