Accuracy Validation of the Cadwell Pulse Oximetry System

Last checked: Nov 20, 2026Latest source record: Jul 25, 2016 (this record)Source: ClinicalTrials.govMethods & limitations

Completed

Status

Completed

Enrollment

11

Actual

Start

Jun 2016

Primary completion

Jun 2016

Study completion Jun 2016

Sponsor

Lead sponsor
Cadwell Industries, Inc.Not on our public-company list
Sponsor type
Industry
Collaborators
Clinimark, LLC
First posted
Jul 25, 2016

More device trials from this sponsor

Design

Study type
Observational
Who can join
All sexes, from 18 years, up to 50 years
Healthy volunteers
Yes

Summary

As written by the sponsor on the registry.

The purpose of this study is to evaluate the SpO2 accuracy and performance of the Cadwell Pulse Oximetry system during non-motion conditions over the range of 70-100% SaO2 as compared to arterial blood samples assessed by CO-Oximetry for SpO2 validation. The end goal is to show the SpO2 accuracy performance of the Cadwell Pulse Oximetry System. It is expected that the Accuracy Root Mean Square (Arms) performance of the Cadwell Pulse Oximetry System will meet a specification of 3 or better in non-motion conditions for the range of 70-100% SaO2 thereby demonstrating an acceptable SpO2 accuracy performance specification.

Conditions

  • Anoxia

Interventions

  • Disposable SensorDevice

    Two adhesive based Pulse Oximeter Probes, S0136J-L were connected to the Cadwell Pulse Oximetry and then placed on each subject to evaluate the SpO2 accuracy performance during steady state non-motion conditions.

  • Reusable SensorDevice

    Two soft reusable Pulse Oximeter Probes, Model S0080D-S were connected to the Cadwell Pulse Oximetry and then placed on each subject to evaluate the SpO2 accuracy performance during steady state non-motion conditions.

  • Control Pulse OximetryDevice

    A Clinimark Control Pulse Oximetry system was also placed on the subject to evaluate the stability of the draws.

Primary outcomes

What the sponsor says the trial is designed to measure.

  1. SpO2 Sensor ValuesTime frame: Through study completion, an average of 1 day

Locations

No sites are listed on the registry yet.

Contact details for sites and study staff are not shown here. To ask about taking part, use the contact information on the registry record.

For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.