All In MedTech

Device Trial Radar · NCT01797705

Revised Algorithm for Next Generation DeVilbiss Healthcare AutoAdjust Continuous Positive Airway Pressure (CPAP)

Revised Algorithm for Next Generation DeVilbiss AutoAdjust CPAP

Last checked: Nov 20, 2026Latest source record: May 29, 2015 (this record)Source: ClinicalTrials.govMethods & limitations

CompletedResults posted on the registry

Status

Completed

Enrollment

33

Actual

Start

Jan 2013

Primary completion

May 2014

Study completion May 2014

Sponsor

Lead sponsor
DeVilbiss Healthcare LLCNot on our public-company list
Sponsor type
Industry
First posted
Feb 22, 2013

More device trials from this sponsor

Design

Study type
Interventional
Phase
Not applicable (typical for device studies)
Allocation
Not applicable (single group)
Model
Single group
Masking
None
Primary purpose
Diagnostic
Who can join
All sexes, from 21 years, up to 75 years
Healthy volunteers
Yes

Summary

As written by the sponsor on the registry.

The purpose of this protocol is to improve the current FDA-approved continuous positive pressure auto-titration algorithm, including treatment precision, residual disease detection characteristics and phenotyping with the DeVilbiss brand AutoAdjust CPAP device.

Conditions

  • Obstructive Sleep Apnea

Interventions

  • DeVilbiss AutoAdjust CPAP with revised algorithmDevice

    Subjects attended a one-night polysomnogram (PSG) titration study. DeVilbiss AutoAdjust CPAP with a revised algorithm was compared to hand-scored PSG to determine whether the machine results were equivalent to hand scores. Each subject acted as their own control for assessing accuracy of sleep event analysis and prescribed CPAP pressure.

Primary outcomes

What the sponsor says the trial is designed to measure.

  1. % Agreement Between Reviewer and Machine Pressure Settings During Sleep StudyTime frame: 1 night

Locations

1 site in 1 country. Site status is shown where the registry gives one.

  • NeuroCare, Inc.

    Newton, Massachusetts

Contact details for sites and study staff are not shown here. To ask about taking part, use the contact information on the registry record.

For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.