All In MedTech

Device Trial Radar · NCT01715948

A Comparison Between Wireless CROS and Bone-anchored Hearing Devices for Single-Sided Deafness: A Pilot Study

Last checked: Nov 19, 2026Latest source record: Apr 5, 2016 (this record)Source: ClinicalTrials.govMethods & limitations

CompletedResults posted on the registry

Status

Completed

Enrollment

9

Actual

Start

Jan 2013

Primary completion

Oct 2013

Study completion Oct 2013

Sponsor

Lead sponsor
Manohar BanceAcademic, hospital or other
Sponsor type
Academic, hospital or other
Collaborators
Dalhousie University
First posted
Oct 29, 2012

More device trials from this sponsor

Design

Study type
Interventional
Phase
Not applicable (typical for device studies)
Allocation
Randomized
Model
Single group
Masking
None
Primary purpose
Treatment
Who can join
All sexes, from 18 years
Healthy volunteers
Yes

Summary

As written by the sponsor on the registry.

People with single-sided deafness (SSD) have significant hearing loss in one ear and normal or near normal hearing in the other ear. Those living with SSD experience several communication difficulties, particularly when listening to speech in the presence of background noise. This problem is worse in situations where the noise is on the side of the good ear and the speech is on the side of the poor ear. The Bone-anchored hearing device (BAHD) and the Contralateral Routing of Signals (CROS) hearing aid are two intervention methods designed to improve hearing in people who have significant hearing loss in one ear only. The BAHD is a surgically implanted device which also includes an external processor; sounds from the poor ear are transmitted to the good ear through skull vibrations. The CROS does not involve surgery; instead a hearing aid is fitted behind each ear, and the sounds on the side of the poor ear are wirelessly transmitted to the good ear. The majority of previous studies comparing the BAHD to the CROS have used older CROS models with basic technology and a wire along the neck to send sounds from the poor ear to the good ear. There is a lack of studies comparing newer digital wireless CROS hearing aids to the BAHD. The current research will compare the effect of the BAHD and CROS hearing aid on speech perception scores when listening to speech in quiet and in noise. The research will also investigate participants' reported benefits with each device during everyday situations. In order to compare the BAHD and CROS, individuals who already have been implanted with a BAHD will be given a trial period with a new wireless digital CROS hearing aid. The research hypothesis states that the CROS hearing aid will be as equally beneficial or greater than the BAHD in improving hearing and participant benefit. If the findings support the hypothesis that the CROS offers just as much benefits as the BAHD, or is an even better alternative, more individuals with SSD may choose to avoid surgery and decide to receive a CROS hearing aid instead.

Conditions

  • Unilateral Hearing Loss
  • Unilateral Deafness

Interventions

  • CROS hearing aidDevice

    Participants were randomly assigned to be fitted with the CROS hearing aid either at the end of the first visit or at the end of the second visit. They were given a 2-week trial with the CROS and asked to refrain from wearing their BAHD during that time. At the end of the first visit, four participants were fitted with the CROS hearing aid and were subsequently tested with this device on their second visit, at which time they were instructed to use the BAHD for the next 2 weeks. This protocol was reversed in the other four participants, with testing after each device experience. Both devices were compared on head shadow effect reduction, speech perception measures in quiet and in noise, self-assessment questionnaires, and daily diaries.

  • Bone-anchored hearing device (BAHD)Device

    All participants had received BAHD implantation prior to the study. They were randomly selected to continue wearing their BAHD for 2 weeks following the end of the first visit or the end of the second visit. Both devices were compared on head shadow effect reduction, speech perception measures in quiet and in noise, self-assessment questionnaires, and daily diaries.

Primary outcomes

What the sponsor says the trial is designed to measure.

  1. Number of 5 Key Words Within 6 Sentence Lists Repeated Correctly in the Presence of Multitalker NoiseTime frame: The QuickSIN unaided was administered at baseline, an average of 2 weeks (with CROS or BAHD) and an average of 4 weeks (with opposite device not previously tested).
  2. Percentage of Words RecognizedTime frame: Word recognition testing (unaided) occurred at baseline, an average of 2 weeks (with CROS or BAHD) and an average of 4 weeks (with opposite device not previously tested)..
  3. Speech Spatial Qualities Questionnaire (SSQ)Time frame: Administered at the end of a 2 week trial with the CROS and at the end of a 2 week use of the BAHD.

Locations

No sites are listed on the registry yet.

Contact details for sites and study staff are not shown here. To ask about taking part, use the contact information on the registry record.

For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.