Device Trial Radar · NCT01679262
Determination of the Optimal Size of Oropharyngeal Airway; Correlation With External Body Measurements
Last checked: Nov 20, 2026Latest source record: Sep 5, 2012 (this record)Source: ClinicalTrials.govMethods & limitations
Status
Completed
Enrollment
149
Actual
Start
May 2010
Primary completion
May 2012
Study completion May 2012
Sponsor
- Lead sponsor
- Yonsei UniversityAcademic, hospital or other
- Sponsor type
- Academic, hospital or other
- First posted
- Sep 5, 2012
Design
- Study type
- Interventional
- Phase
- Not applicable (typical for device studies)
- Allocation
- Not applicable (single group)
- Model
- Single group
- Masking
- Single (outcomes assessor)
- Primary purpose
- Prevention
- Who can join
- All sexes, from 20 years, up to 70 years
- Healthy volunteers
- No
Summary
As written by the sponsor on the registry.
There are several studies that have measured various oropharyngeal airway space distances using radiological method. However, direct confirmation of optimal size of oropharyneal airways (OPAs) in individual using fiberoptic bronchoscope (FOB) is lacking as now. The purposes of this study were to directly determinate the proper size of OPA in individual by FOB via different sizes of OPA in adults under general anestheisia. Additionally, we examine the relationships among the size of OPA, straight length from the incisors to the tip of the epiglottis by laryngoscopy, patient height, and external facial measurements.
Conditions
- Ear,Nose and Throat Surgery
Interventions
Guedel type color-coded OPAsDevice
The OPAs was used four different sizes as No. 8 (80 mm, green), 9 (90 mm, yellow), 10 (100 mm, red), and 11 (110 mm, orange) in regular sequence. The curvilinear distance from the incisors to the tip of the epiglottis was measured by FOB (Olympus LF-GP; Olympus Optical Co., Tokyo, Japan) with tape marking via each size of OPAs. The distance from the distal end of OPA to the tip of the epiglottis was calculated by the difference between measured distance from the incisors to the tip of the epiglottis via each OPA by FOB and typical distance from the flange to distal end of each OPA via inside tube of OPA. When the size of OPAs was longer than distance from the incisors to the tip of the epiglottis, OPAs were partly withdrawn and repositioned the distal end of OPAs is placed just above the tip of epiglottis. Then, the length from the incisors to the flange of OPAs was measured. .
Primary outcomes
What the sponsor says the trial is designed to measure.
- Optimal size of OPATime frame: 1 min after each FOB measurement
Locations
1 site in 1 country. Site status is shown where the registry gives one.
Wyun Kon Park, Yonsei University College of Medicine, Department of Anesthesiology and Pain Medicine
Seoul, Seoul, South Korea
Contact details for sites and study staff are not shown here. To ask about taking part, use the contact information on the registry record.
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.