Device Trial Radar · NCT00575822
A Study to Determine the Effectiveness of Endoscopic Full-Thickness Plication for the Treatment of GERD
Last checked: Nov 20, 2026Latest source record: Jul 16, 2008 (this record)Source: ClinicalTrials.govMethods & limitations
Sponsor
- Lead sponsor
- NDO Surgical, Inc.Not on our public-company list
- Sponsor type
- Industry
- First posted
- Dec 18, 2007
Design
- Study type
- Interventional
- Phase
- Not applicable (typical for device studies)
- Allocation
- Randomized
- Model
- Crossover
- Masking
- Double (participant, outcomes assessor)
- Primary purpose
- Treatment
- Who can join
- All sexes, from 18 years
- Healthy volunteers
- No
Summary
As written by the sponsor on the registry.
The aim of this study was to determine the effectiveness of endoscopic full-thickness plication (Plicator; NDO Surgical, Inc., Mansfield, MA) for the treatment of GERD in comparison to a sham procedure. Patients with symptomatic GERD requiring maintenance proton pump inhibitor therapy were entered into a randomized, single-blind, prospective, multicenter trial. Seventy-eight patients were randomly assigned to undergo endoscopic full-thickness restructuring of the gastric cardia with transmural suture. Eighty-one patients underwent a sham procedure. Group assignments were revealed following the 3-month evaluation. The primary endpoint was ≥ 50% improvement in GERD-HRQL score. Secondary endpoints included medication use and esophageal acid exposure. Patients achieving ≥ 50% improvement in GERD-HRQL score at 3-months versus baseline off-meds were considered responsive to their assigned procedure. Patients who failed to reach this level of improvement at 3-months were considered non-responders. Analysis of these dichotomized variables (responder/non-responder) was done using Fisher's exact test comparing the proportion of responders between the active and sham groups. Intent-to-treat analysis was also performed. The null hypothesis was that the proportion of responders was the same in both the active and sham groups. Testing was done at the 5% level of significance (alpha=0.05).
Conditions
- Gastroesophageal Reflux Disease (GERD)
Interventions
NDO Full-thickness PlicatorDevice
A low-profile (≤6mm) gastroscope was advanced and an EGD followed by Savary wire placement was performed. The gastroscope was then removed and the Plicator was advanced directly over the wire and into the stomach. The gastroscope was re-introduced through the Plicator, and the Plicator was retroflexed under direct endoscopic visualization to the anterior gastric cardia, approximately 1cm below the gastroesophageal (GE) junction. The Plicator arms were opened and the endoscopic tissue retractor was advanced deeply into the gastric cardia. The gastric wall was then retracted into the open arms of the Plicator. The arms were closed and a single, transmural pledgeted suture was deployed. The Plicator and gastroscope were then removed and the gastroscope re-inserted to evaluate the post-plication GE junction.
Sham ControlDevice
The sham procedure was identical to the treatment procedure through the positioning of the retroflexed Plicator 1cm below the GE junction. The Plicator instrument remained in this position for 15 minutes, based on average procedure times established in previous studies. During the sham procedure, physician investigators were required to talk through the procedure steps to maintain the patient blind.
Primary outcomes
What the sponsor says the trial is designed to measure.
- >/=50% improvement in GERD-Health Related Quality of Life ScoreTime frame: 3-months post-procedure
Locations
15 sites in 3 countries. Site status is shown where the registry gives one.
University of California at Irvine Medical Center
Orange, California
14 more study sites with a free account
Contact details for sites and study staff are not shown here. To ask about taking part, use the contact information on the registry record.
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.