Device type · OEV
Cardiopulmonary Resuscitation Aid Kit
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
OEV
Anesthesiology
Device class
Class II
FDA decisions since 2015
0
510(k), De Novo and PMA
Adverse event reports
22
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Cardiopulmonary Resuscitation Aid Kit
- Product code
- OEV
- Device class
- Class II
- Review panel
- Anesthesiology
- Regulation
- 21 CFR 868.5915
- Specialty
- Anesthesiology & Respiratory
- FDA record
- Classification database entry for OEV
Latest FDA decisions
No 510(k), De Novo or PMA decisions for this product code since 2015.
Reports received
22
Received Nov 1, 2025 to Nov 19, 2026
Malfunction reports
12
55% of reports
Injury reports
6
27% of reports
Death reports
4
18% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 0 | 0 | 0 | 0 |
| Jan 24 | 0 | 1 | 0 | 0 |
| Feb 24 | 0 | 0 | 0 | 0 |
| Mar 24 | 1 | 0 | 0 | 0 |
| Apr 24 | 2 | 0 | 2 | 0 |
| May 24 | 2 | 0 | 0 | 0 |
| Jun 24 | 2 | 0 | 0 | 0 |
| Jul 24 | 1 | 0 | 2 | 0 |
| Aug 24 | 2 | 0 | 0 | 0 |
| Sep 24 | 0 | 0 | 2 | 0 |
| Oct 24 | 0 | 0 | 0 | 0 |
| Nov 24 | 1 | 1 | 0 | 0 |
| Dec 24 | 1 | 3 | 0 | 0 |
| Jan 25 | 3 | 0 | 0 | 0 |
| Feb 25 | 1 | 0 | 0 | 0 |
| Mar 25 | 0 | 0 | 0 | 0 |
| Apr 25 | 1 | 1 | 2 | 0 |
| May 25 | 3 | 3 | 0 | 0 |
| Jun 25 | 8 | 0 | 0 | 0 |
| Jul 25 | 2 | 0 | 2 | 0 |
| Aug 25 | 3 | 0 | 0 | 0 |
| Sep 25 | 1 | 0 | 0 | 0 |
| Oct 25 | 3 | 0 | 0 | 0 |
| Nov 25 | 0 | 2 | 0 | 0 |
| Dec 25 | 0 | 0 | 0 | 0 |
| Jan 26 | 1 | 0 | 0 | 0 |
| Feb 26 | 2 | 0 | 0 | 0 |
| Mar 26 | 0 | 0 | 0 | 0 |
| Apr 26 | 2 | 0 | 0 | 0 |
| May 26 | 3 | 2 | 0 | 0 |
| Jun 26 | 2 | 0 | 2 | 0 |
| Jul 26 | 0 | 2 | 0 | 0 |
| Aug 26 | 0 | 0 | 0 | 0 |
| Sep 26 | 0 | 0 | 0 | 0 |
| Oct 26 | 2 | 0 | 2 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
- Ambu21
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Component Missing10
- Gas/Air Leak3
- Detachment of Device or Device Component2
- Device Handling Problem2
- Improper or Incorrect Procedure or Method2
- Inadequate User Interface2
- Inflation Problem2
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions11
- Hypoxia4
- Brain Injury2
- Cyanosis2
- Insufficient Information2
- Asystole1
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
- Respirator Crash Cart KitTerminated
Z-2466-2020 · Started May 19, 2020
Reason given: Roi CPS, LLC has become aware of a labeling error that affects select custom procedure kits. The kits are labeled with an expiration date that is later than that of the soonest kit component to expire in the kits.
Cause as FDA categorized it: Error in labeling
Z-1547-2018 · Started Apr 27, 2017 · Posted by FDA May 10, 2018
Reason given: The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
Cause as FDA categorized it: Component design/selection
Z-1548-2018 · Started Apr 27, 2017 · Posted by FDA May 10, 2018
Reason given: The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
Cause as FDA categorized it: Component design/selection
Z-1549-2018 · Started Apr 27, 2017 · Posted by FDA May 10, 2018
Reason given: The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
Cause as FDA categorized it: Component design/selection
Z-1550-2018 · Started Apr 27, 2017 · Posted by FDA May 10, 2018
Reason given: The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
Cause as FDA categorized it: Component design/selection
Z-1551-2018 · Started Apr 27, 2017 · Posted by FDA May 10, 2018
Reason given: The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
Cause as FDA categorized it: Component design/selection
Z-1552-2018 · Started Apr 27, 2017 · Posted by FDA May 10, 2018
Reason given: The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
Cause as FDA categorized it: Component design/selection
Z-1553-2018 · Started Apr 27, 2017 · Posted by FDA May 10, 2018
Reason given: The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
Cause as FDA categorized it: Component design/selection
Showing the 8 newest of 36 recall records in FDA's data.
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.