Device type · KNH
Laparoscopic Contraceptive Tubal Occlusion Device
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
KNH
Obstetrics & Gynecology
Device class
Class III
FDA decisions since 2015
0
510(k), De Novo and PMA
Adverse event reports
9
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Laparoscopic Contraceptive Tubal Occlusion Device
- Product code
- KNH
- Device class
- Class III
- Review panel
- Obstetrics & Gynecology
- Regulation
- 21 CFR 884.5380
- Specialty
- Obstetrics & Gynecology
- FDA record
- Classification database entry for KNH
Latest FDA decisions
No 510(k), De Novo or PMA decisions for this product code since 2015.
Reports received
9
Received Nov 1, 2025 to Nov 19, 2026. Includes 1 of another or unstated type.
Malfunction reports
0
0.0% of reports
Injury reports
8
89% of reports
Death reports
0
0.0% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 0 | 0 | 0 | 0 |
| Jan 24 | 0 | 1 | 0 | 0 |
| Feb 24 | 0 | 0 | 0 | 0 |
| Mar 24 | 0 | 0 | 0 | 0 |
| Apr 24 | 0 | 1 | 0 | 0 |
| May 24 | 0 | 4 | 0 | 0 |
| Jun 24 | 0 | 0 | 0 | 0 |
| Jul 24 | 0 | 2 | 0 | 0 |
| Aug 24 | 0 | 1 | 0 | 0 |
| Sep 24 | 0 | 0 | 0 | 0 |
| Oct 24 | 0 | 0 | 0 | 0 |
| Nov 24 | 0 | 0 | 0 | 0 |
| Dec 24 | 0 | 0 | 0 | 0 |
| Jan 25 | 0 | 1 | 0 | 0 |
| Feb 25 | 0 | 1 | 0 | 0 |
| Mar 25 | 0 | 1 | 0 | 0 |
| Apr 25 | 0 | 0 | 0 | 0 |
| May 25 | 0 | 0 | 0 | 0 |
| Jun 25 | 0 | 0 | 0 | 0 |
| Jul 25 | 0 | 0 | 0 | 0 |
| Aug 25 | 0 | 1 | 0 | 0 |
| Sep 25 | 0 | 0 | 0 | 0 |
| Oct 25 | 0 | 0 | 0 | 0 |
| Nov 25 | 0 | 2 | 0 | 0 |
| Dec 25 | 0 | 1 | 0 | 0 |
| Jan 26 | 0 | 0 | 0 | 0 |
| Feb 26 | 0 | 0 | 0 | 0 |
| Mar 26 | 0 | 0 | 0 | 0 |
| Apr 26 | 0 | 0 | 0 | 0 |
| May 26 | 0 | 0 | 0 | 0 |
| Jun 26 | 0 | 5 | 0 | 0 |
| Jul 26 | 0 | 0 | 0 | 0 |
| Aug 26 | 0 | 0 | 0 | 0 |
| Sep 26 | 0 | 0 | 0 | 0 |
| Oct 26 | 0 | 0 | 0 | 1 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Migration7
- Device Dislodged or Dislocated4
- Entrapment of Device4
2 more in this ranking with a free account
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Abdominal Pain5
- Emotional Changes4
- Foreign Body Reaction4
5 more in this ranking with a free account
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-0419-2015 · Started Oct 21, 2014 · Posted by FDA Jun 26, 2015
Reason given: All packages of Falope Ring Band Applicator kits are being recalled due to compromises in sterile packaging that could be associated with bacterial contamination that might lead to patient infection.
Cause as FDA categorized it: Process control
Richard Wolf Medical Instruments Corp.
Z-0037-2015 · Started Jul 9, 2014 · Posted by FDA Nov 6, 2014
Reason given: Richard Wolf Medical Instrument Corporation is recalling Hulka Clip Tubal Occluding Device lot 51000232. The devices are being recalled as a part of an investigation into a biocompatibility test discrepancy where the recalled lot (lot 51000232) passed final release testing, but the subsequent lot failed...
Cause as FDA categorized it: Nonconforming Material/Component
Z-2269-2014 · Started Jul 8, 2014 · Posted by FDA Aug 19, 2014
Reason given: Falope-Ring Dilator for Reusable Falope-Ring Band Applicators have been sold as a stand-alone non-sterile component. The dilator is intended to assist placement of the Falope-Ring band on the reusable Falope applicator.
Cause as FDA categorized it: Packaging process control
1 more recall records with a free account
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.