Device type · KNC
Table, Obstetric (And Accessories)
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
KNC
Obstetrics & Gynecology
Device class
Class II
FDA decisions since 2015
0
510(k), De Novo and PMA
Adverse event reports
154
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Table, Obstetric (And Accessories)
- Product code
- KNC
- Device class
- Class II
- Review panel
- Obstetrics & Gynecology
- Regulation
- 21 CFR 884.4900
- Specialty
- Obstetrics & Gynecology
- FDA record
- Classification database entry for KNC
Latest FDA decisions
No 510(k), De Novo or PMA decisions for this product code since 2015.
Reports received
154
Received Nov 1, 2025 to Nov 19, 2026
Malfunction reports
151
98% of reports
Injury reports
3
1.9% of reports
Death reports
0
0.0% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 3 | 0 | 0 | 0 |
| Jan 24 | 2 | 0 | 0 | 0 |
| Feb 24 | 2 | 0 | 0 | 0 |
| Mar 24 | 3 | 0 | 0 | 0 |
| Apr 24 | 3 | 0 | 0 | 0 |
| May 24 | 8 | 0 | 0 | 0 |
| Jun 24 | 4 | 0 | 0 | 0 |
| Jul 24 | 5 | 0 | 0 | 0 |
| Aug 24 | 2 | 0 | 0 | 0 |
| Sep 24 | 5 | 0 | 0 | 0 |
| Oct 24 | 5 | 0 | 0 | 0 |
| Nov 24 | 17 | 0 | 0 | 0 |
| Dec 24 | 14 | 1 | 0 | 0 |
| Jan 25 | 20 | 0 | 0 | 0 |
| Feb 25 | 8 | 0 | 0 | 0 |
| Mar 25 | 8 | 0 | 0 | 0 |
| Apr 25 | 9 | 0 | 0 | 0 |
| May 25 | 9 | 0 | 0 | 0 |
| Jun 25 | 10 | 0 | 0 | 0 |
| Jul 25 | 9 | 0 | 0 | 0 |
| Aug 25 | 4 | 0 | 0 | 0 |
| Sep 25 | 8 | 0 | 0 | 0 |
| Oct 25 | 9 | 0 | 0 | 0 |
| Nov 25 | 12 | 0 | 0 | 0 |
| Dec 25 | 8 | 0 | 0 | 0 |
| Jan 26 | 3 | 0 | 0 | 0 |
| Feb 26 | 9 | 0 | 0 | 0 |
| Mar 26 | 21 | 0 | 0 | 0 |
| Apr 26 | 11 | 1 | 0 | 0 |
| May 26 | 14 | 0 | 0 | 0 |
| Jun 26 | 20 | 1 | 0 | 0 |
| Jul 26 | 11 | 0 | 0 | 0 |
| Aug 26 | 15 | 1 | 0 | 0 |
| Sep 26 | 11 | 0 | 0 | 0 |
| Oct 26 | 16 | 0 | 0 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
- Baxter154
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Unintended Movement115
- Material Split, Cut or Torn16
- Activation Problem11
7 more in this ranking with a free account
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions151
- Insufficient Information2
- Ambulation or Postural Difficulties1
6 more in this ranking with a free account
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-1816-2025 · Started May 6, 2025 · Posted by FDA May 23, 2025
Reason given: There is a potential for improper crimps on the signal wires of the actuators which could result in a loss of ability to lower the bed or the foot section of the bed.
Cause as FDA categorized it: Nonconforming Material/Component
Z-2156-2021 · Started May 28, 2021
Reason given: Latch mechanism on the Lift-Off foot section of the Affinity Four Birthing Bed is damaged, it could potentially cause the installed foot section to be improperly engaged onto the bed, could result in injury to the user due to a fall.
Cause as FDA categorized it: Device Design
Z-2750-2016 · Started Jul 15, 2016 · Posted by FDA Sep 8, 2016
Cause as FDA categorized it: Nonconforming Material/Component
9 more recall records with a free account
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.