Device type · JDZ
Mixer, Cement, For Clinical Use
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
JDZ
Orthopedic
Device class
Class I
FDA decisions since 2015
0
510(k), De Novo and PMA
Adverse event reports
238
Received Nov 1, 2025 to Nov 20, 2026
How FDA classifies it
- Device type
- Mixer, Cement, For Clinical Use
- Product code
- JDZ
- Device class
- Class I
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.4210
- Specialty
- Orthopedic Surgery
- FDA record
- Classification database entry for JDZ
Latest FDA decisions
No 510(k), De Novo or PMA decisions for this product code since 2015.
Reports received
238
Received Nov 1, 2025 to Nov 20, 2026
Malfunction reports
234
98% of reports
Injury reports
4
1.7% of reports
Death reports
0
0.0% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 2 | 0 | 0 | 0 |
| Jan 24 | 4 | 0 | 0 | 0 |
| Feb 24 | 1 | 1 | 0 | 0 |
| Mar 24 | 2 | 0 | 0 | 0 |
| Apr 24 | 4 | 0 | 0 | 0 |
| May 24 | 0 | 0 | 0 | 0 |
| Jun 24 | 0 | 0 | 0 | 0 |
| Jul 24 | 5 | 0 | 0 | 0 |
| Aug 24 | 0 | 0 | 0 | 0 |
| Sep 24 | 0 | 0 | 0 | 0 |
| Oct 24 | 11 | 0 | 0 | 0 |
| Nov 24 | 0 | 0 | 0 | 0 |
| Dec 24 | 0 | 0 | 0 | 0 |
| Jan 25 | 7 | 0 | 0 | 0 |
| Feb 25 | 0 | 0 | 0 | 0 |
| Mar 25 | 2 | 0 | 0 | 0 |
| Apr 25 | 4 | 0 | 0 | 0 |
| May 25 | 0 | 0 | 0 | 0 |
| Jun 25 | 0 | 0 | 0 | 0 |
| Jul 25 | 1 | 0 | 0 | 0 |
| Aug 25 | 2 | 0 | 0 | 0 |
| Sep 25 | 1 | 0 | 0 | 0 |
| Oct 25 | 2 | 0 | 0 | 0 |
| Nov 25 | 0 | 0 | 0 | 0 |
| Dec 25 | 1 | 0 | 0 | 0 |
| Jan 26 | 5 | 0 | 0 | 0 |
| Feb 26 | 6 | 0 | 0 | 0 |
| Mar 26 | 86 | 0 | 0 | 0 |
| Apr 26 | 106 | 0 | 0 | 0 |
| May 26 | 21 | 1 | 0 | 0 |
| Jun 26 | 1 | 1 | 0 | 0 |
| Jul 26 | 1 | 0 | 0 | 0 |
| Aug 26 | 4 | 1 | 0 | 0 |
| Sep 26 | 0 | 1 | 0 | 0 |
| Oct 26 | 3 | 0 | 0 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
- Stryker13
2 more in this ranking with a free account
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Unsealed Device Packaging215
- Detachment of Device or Device Component11
- Delivered as Unsterile Product3
7 more in this ranking with a free account
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions233
- Insufficient Information2
- Medical device site infection1
2 more in this ranking with a free account
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-1946-2026 · Started Mar 30, 2026 · Posted by FDA Apr 24, 2026
Reason given: Ten complaints have been received identifying issues at the time of use related to the package seal, including incomplete seals, wrinkles in the seals or peeling seals. Compromised sterile barrier may lead to clinically insignificant extension of surgery or infection leading to surgical intervention.
Cause as FDA categorized it: Vendor change control
- Heraeus, PALAMIX uno. Material Number: 66057893.Open, Classified
Heraeus Medical GmbH (Dental Division)
Z-1997-2026 · Started Mar 25, 2026 · Posted by FDA Apr 29, 2026
Reason given: Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).
Cause as FDA categorized it: Under Investigation by firm
- Heraeus, PALAMIX duo. Material Number: 66057897.Open, Classified
Heraeus Medical GmbH (Dental Division)
Z-1998-2026 · Started Mar 25, 2026 · Posted by FDA Apr 29, 2026
Reason given: Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).
Cause as FDA categorized it: Under Investigation by firm
15 more recall records with a free account
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.