Device type · IDP
Microtome, Cryostat
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
IDP
Pathology
Device class
Class I
FDA decisions since 2015
0
510(k), De Novo and PMA
Adverse event reports
2
Received Nov 1, 2025 to Nov 20, 2026
How FDA classifies it
- Device type
- Microtome, Cryostat
- Product code
- IDP
- Device class
- Class I
- Review panel
- Pathology
- Regulation
- 21 CFR 864.3010
- Specialty
- Laboratory Medicine & Pathology
- FDA record
- Classification database entry for IDP
Latest FDA decisions
No 510(k), De Novo or PMA decisions for this product code since 2015.
Reports received
2
Received Nov 1, 2025 to Nov 20, 2026
Malfunction reports
0
0.0% of reports
Injury reports
2
100% of reports
Death reports
0
0.0% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 0 | 1 | 0 | 0 |
| Jan 24 | 0 | 2 | 0 | 0 |
| Feb 24 | 0 | 0 | 0 | 0 |
| Mar 24 | 0 | 0 | 0 | 0 |
| Apr 24 | 0 | 0 | 0 | 0 |
| May 24 | 0 | 0 | 0 | 0 |
| Jun 24 | 0 | 0 | 0 | 0 |
| Jul 24 | 0 | 0 | 0 | 0 |
| Aug 24 | 0 | 1 | 0 | 0 |
| Sep 24 | 0 | 0 | 0 | 0 |
| Oct 24 | 0 | 0 | 0 | 0 |
| Nov 24 | 0 | 0 | 0 | 0 |
| Dec 24 | 0 | 0 | 0 | 0 |
| Jan 25 | 0 | 0 | 0 | 0 |
| Feb 25 | 0 | 0 | 0 | 0 |
| Mar 25 | 0 | 0 | 0 | 0 |
| Apr 25 | 0 | 0 | 0 | 0 |
| May 25 | 0 | 0 | 0 | 0 |
| Jun 25 | 0 | 1 | 0 | 0 |
| Jul 25 | 0 | 0 | 0 | 0 |
| Aug 25 | 0 | 0 | 0 | 0 |
| Sep 25 | 0 | 0 | 0 | 0 |
| Oct 25 | 0 | 0 | 0 | 0 |
| Nov 25 | 0 | 0 | 0 | 0 |
| Dec 25 | 0 | 0 | 0 | 0 |
| Jan 26 | 0 | 0 | 0 | 0 |
| Feb 26 | 0 | 0 | 0 | 0 |
| Mar 26 | 0 | 2 | 0 | 0 |
| Apr 26 | 0 | 0 | 0 | 0 |
| May 26 | 0 | 0 | 0 | 0 |
| Jun 26 | 0 | 0 | 0 | 0 |
| Jul 26 | 0 | 0 | 0 | 0 |
| Aug 26 | 0 | 0 | 0 | 0 |
| Sep 26 | 0 | 0 | 0 | 0 |
| Oct 26 | 0 | 0 | 0 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Inadequate or Insufficient Training2
- Improper or Incorrect Procedure or Method1
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Laceration(s)2
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
- Brand Name: Leica Biosystems Cryostats Product Name: CM1950 Model/Catalog Number: 1491950C4US Software Version: firmware V3.01Open, Classified
LEICA BIOSYSTEMS NUSSLOCH GMBH
Z-1247-2025 · Started Jan 30, 2025 · Posted by FDA Feb 27, 2025
Reason given: Firm became aware of software issues related to most recent firmware update which could result in loss of patient tissue.
Cause as FDA categorized it: Under Investigation by firm
- Leica Biosystems Cryostat Model Number CM1100, Material Number 14046931130, for freezing and sectioning tissue samplesOpen, Classified
LEICA BIOSYSTEMS NUSSLOCH GMBH
Z-0085-2025 · Started Sep 4, 2024 · Posted by FDA Oct 15, 2024
Reason given: As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat...
Cause as FDA categorized it: Labeling design
- Leica Biosystems Cryostat Model Number CM1510 S, Material Numbers: a) 491510C002, b) 1491510C003; for freezing and sectioning tissue samplesOpen, Classified
LEICA BIOSYSTEMS NUSSLOCH GMBH
Z-0086-2025 · Started Sep 4, 2024 · Posted by FDA Oct 15, 2024
Reason given: As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat...
Cause as FDA categorized it: Labeling design
7 more recall records with a free account
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.