Device type · FWZ
Operating Room Accessories Table Tray
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
FWZ
General & Plastic Surgery
Device class
Class I
FDA decisions since 2015
0
510(k), De Novo and PMA
Adverse event reports
82
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Operating Room Accessories Table Tray
- Product code
- FWZ
- Device class
- Class I
- Review panel
- General & Plastic Surgery
- Regulation
- 21 CFR 878.4950
- Specialty
- General & Plastic Surgery
- FDA record
- Classification database entry for FWZ
Latest FDA decisions
No 510(k), De Novo or PMA decisions for this product code since 2015.
Reports received
82
Received Nov 1, 2025 to Nov 19, 2026. Includes 2 of another or unstated type.
Malfunction reports
69
84% of reports
Injury reports
11
13% of reports
Death reports
0
0.0% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 2 | 0 | 0 | 0 |
| Jan 24 | 4 | 3 | 0 | 0 |
| Feb 24 | 1 | 1 | 0 | 0 |
| Mar 24 | 2 | 3 | 0 | 0 |
| Apr 24 | 3 | 2 | 0 | 0 |
| May 24 | 8 | 0 | 0 | 0 |
| Jun 24 | 9 | 0 | 0 | 0 |
| Jul 24 | 0 | 0 | 0 | 0 |
| Aug 24 | 6 | 0 | 0 | 0 |
| Sep 24 | 1 | 5 | 0 | 0 |
| Oct 24 | 2 | 0 | 0 | 0 |
| Nov 24 | 3 | 4 | 0 | 0 |
| Dec 24 | 9 | 2 | 0 | 0 |
| Jan 25 | 3 | 0 | 0 | 0 |
| Feb 25 | 5 | 3 | 0 | 0 |
| Mar 25 | 3 | 2 | 0 | 0 |
| Apr 25 | 2 | 0 | 0 | 0 |
| May 25 | 4 | 0 | 0 | 0 |
| Jun 25 | 4 | 1 | 0 | 0 |
| Jul 25 | 1 | 3 | 0 | 0 |
| Aug 25 | 0 | 5 | 0 | 0 |
| Sep 25 | 3 | 0 | 0 | 0 |
| Oct 25 | 5 | 0 | 0 | 0 |
| Nov 25 | 4 | 1 | 0 | 0 |
| Dec 25 | 2 | 1 | 0 | 0 |
| Jan 26 | 7 | 0 | 0 | 0 |
| Feb 26 | 4 | 5 | 0 | 0 |
| Mar 26 | 5 | 2 | 0 | 0 |
| Apr 26 | 2 | 0 | 0 | 0 |
| May 26 | 3 | 0 | 0 | 0 |
| Jun 26 | 2 | 0 | 0 | 0 |
| Jul 26 | 5 | 2 | 0 | 0 |
| Aug 26 | 16 | 0 | 0 | 0 |
| Sep 26 | 12 | 0 | 0 | 1 |
| Oct 26 | 7 | 0 | 0 | 1 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
- Baxter31
4 more in this ranking with a free account
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Device Slipped17
- Unstable15
- Unintended Movement8
7 more in this ranking with a free account
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions66
- Laceration(s)9
- Insufficient Information4
2 more in this ranking with a free account
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-1289-2024 · Started Feb 14, 2024 · Posted by FDA Mar 12, 2024
Reason given: Three lots of the Boot Assembly, Traction Device are being recalled due to due to a customer report of an incorrectly oriented locking pin which caused the boot to detach from the traction system. If an impacted device is used during surgery, the boot may detach from the traction system, potentially leading to patient...
Cause as FDA categorized it: Under Investigation by firm
Z-1867-2017 · Started Mar 28, 2017 · Posted by FDA Apr 19, 2017
Reason given: Mizuho has discovered that the 5319-37 Wilson Frame Pads have a design deficiency in the construction of their mounting sleeves that may allow the patient to contact the underlying carbon fiber structure of the Wilson frame and could lead to a pressure injury.
Cause as FDA categorized it: Process control
Maquet Cardiovascular Us Sales, Llc
Z-0247-2016 · Started Aug 17, 2015 · Posted by FDA Nov 9, 2015
Reason given: Two serious thromboembolic events were reported involving the Adjustable Positioning Elements (Part Number 1007.28A0) in conjunction with Universal Frame (Part Number 1007.24F0) for lumbar spine surgery. The hip elements may have a negative effect on patients- venous backflow under certain conditions and may have been...
Cause as FDA categorized it: Under Investigation by firm
34 more recall records with a free account
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.