Device type · FSA
Lift, Patient, Non-Ac-Powered
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
FSA
General Hospital
Device class
Class I
FDA decisions since 2015
0
510(k), De Novo and PMA
Adverse event reports
2,289
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Lift, Patient, Non-Ac-Powered
- Product code
- FSA
- Device class
- Class I
- Review panel
- General Hospital
- Regulation
- 21 CFR 880.5510
- Specialty
- Hospital Medicine & Nursing
- FDA record
- Classification database entry for FSA
Latest FDA decisions
No 510(k), De Novo or PMA decisions for this product code since 2015.
Reports received
2,289
Received Nov 1, 2025 to Nov 19, 2026. Includes 14 of another or unstated type.
Malfunction reports
2,150
94% of reports
Injury reports
99
4.3% of reports
Death reports
26
1.1% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 23 | 6 | 0 | 0 |
| Jan 24 | 13 | 12 | 3 | 0 |
| Feb 24 | 17 | 5 | 0 | 0 |
| Mar 24 | 13 | 10 | 5 | 0 |
| Apr 24 | 21 | 7 | 1 | 0 |
| May 24 | 17 | 14 | 3 | 0 |
| Jun 24 | 16 | 14 | 1 | 0 |
| Jul 24 | 26 | 8 | 2 | 0 |
| Aug 24 | 18 | 8 | 3 | 0 |
| Sep 24 | 13 | 12 | 0 | 0 |
| Oct 24 | 21 | 4 | 0 | 0 |
| Nov 24 | 38 | 7 | 3 | 0 |
| Dec 24 | 27 | 9 | 0 | 0 |
| Jan 25 | 26 | 10 | 3 | 0 |
| Feb 25 | 45 | 12 | 3 | 0 |
| Mar 25 | 28 | 11 | 2 | 0 |
| Apr 25 | 19 | 15 | 1 | 0 |
| May 25 | 17 | 7 | 2 | 0 |
| Jun 25 | 17 | 4 | 4 | 0 |
| Jul 25 | 30 | 12 | 0 | 0 |
| Aug 25 | 19 | 14 | 4 | 0 |
| Sep 25 | 29 | 12 | 0 | 0 |
| Oct 25 | 15 | 16 | 1 | 0 |
| Nov 25 | 23 | 9 | 0 | 0 |
| Dec 25 | 20 | 8 | 0 | 0 |
| Jan 26 | 18 | 13 | 2 | 0 |
| Feb 26 | 22 | 5 | 3 | 0 |
| Mar 26 | 19 | 6 | 4 | 0 |
| Apr 26 | 38 | 11 | 0 | 0 |
| May 26 | 54 | 7 | 2 | 0 |
| Jun 26 | 45 | 6 | 2 | 0 |
| Jul 26 | 704 | 9 | 6 | 1 |
| Aug 26 | 467 | 9 | 2 | 4 |
| Sep 26 | 590 | 7 | 4 | 6 |
| Oct 26 | 150 | 9 | 1 | 3 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
- Medline Industries1,853
- Baxter159
7 more in this ranking with a free account
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Electrical /Electronic Property Problem877
- Fracture219
- Battery Problem148
7 more in this ranking with a free account
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions2,042
- Fall141
- Bone Fracture(s)46
7 more in this ranking with a free account
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
ARJOHUNTLEIGH POLSKA Sp. z.o.o.
Z-0315-2027 · Started Sep 25, 2026 · Posted by FDA Oct 29, 2026
Reason given: The side support of the shower trolley may open unintentionally when the handles are not fully locked and additional force is applied during patient handling. If the side support is not fully locked and an additional force is applied by the caregiver on the handle (e.g., during turning or repositioning), the panel may...
Cause as FDA categorized it: Under Investigation by firm
Z-0151-2027 · Started Sep 14, 2026 · Posted by FDA Oct 13, 2026
Reason given: Potential issue involving motor shaft breakage combined with reduced performance of the Single Fault Safety (SFS) system may lead to instances where the lift strap lowers more quickly than expected during use. Patients are at some risk for injury if a significant fall/drop were to occur due to this issue.
Cause as FDA categorized it: Process control
ARJOHUNTLEIGH POLSKA Sp. z.o.o.
Z-1811-2026 · Started Mar 10, 2026 · Posted by FDA Apr 10, 2026
Reason given: An internal component of the actuators in a certain production batch may not meet the required mechanical strength. This could result, in the worst case scenario, in a sudden loss of the actuator's ability to hold the load, leading to rapid and uncontrolled downward movement of the lifting arm.
Cause as FDA categorized it: Process control
140 more recall records with a free account
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.