Device type · EAP
Unit, X-Ray, Intraoral
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
EAP
Radiology
Device class
Class II
FDA decisions since 2015
0
510(k), De Novo and PMA
Adverse event reports
342
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Unit, X-Ray, Intraoral
- Product code
- EAP
- Device class
- Class II
- Review panel
- Radiology
- Regulation
- 21 CFR 872.1810
- Specialty
- Radiology & Imaging
- FDA record
- Classification database entry for EAP
Latest FDA decisions
No 510(k), De Novo or PMA decisions for this product code since 2015.
Reports received
342
Received Nov 1, 2025 to Nov 19, 2026
Malfunction reports
342
100% of reports
Injury reports
0
0.0% of reports
Death reports
0
0.0% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 0 | 0 | 0 | 0 |
| Jan 24 | 0 | 0 | 0 | 0 |
| Feb 24 | 0 | 0 | 0 | 0 |
| Mar 24 | 0 | 0 | 0 | 0 |
| Apr 24 | 0 | 0 | 0 | 0 |
| May 24 | 0 | 0 | 0 | 0 |
| Jun 24 | 0 | 0 | 0 | 0 |
| Jul 24 | 0 | 0 | 0 | 0 |
| Aug 24 | 0 | 0 | 0 | 0 |
| Sep 24 | 0 | 0 | 0 | 0 |
| Oct 24 | 0 | 0 | 0 | 0 |
| Nov 24 | 0 | 0 | 0 | 0 |
| Dec 24 | 0 | 0 | 0 | 0 |
| Jan 25 | 0 | 0 | 0 | 0 |
| Feb 25 | 0 | 0 | 0 | 0 |
| Mar 25 | 1 | 0 | 0 | 0 |
| Apr 25 | 0 | 0 | 0 | 0 |
| May 25 | 0 | 0 | 0 | 0 |
| Jun 25 | 0 | 0 | 0 | 0 |
| Jul 25 | 0 | 0 | 0 | 0 |
| Aug 25 | 0 | 0 | 0 | 0 |
| Sep 25 | 0 | 0 | 0 | 0 |
| Oct 25 | 0 | 0 | 0 | 0 |
| Nov 25 | 0 | 0 | 0 | 0 |
| Dec 25 | 0 | 0 | 0 | 0 |
| Jan 26 | 0 | 0 | 0 | 0 |
| Feb 26 | 63 | 0 | 0 | 0 |
| Mar 26 | 35 | 0 | 0 | 0 |
| Apr 26 | 69 | 0 | 0 | 0 |
| May 26 | 41 | 0 | 0 | 0 |
| Jun 26 | 51 | 0 | 0 | 0 |
| Jul 26 | 47 | 0 | 0 | 0 |
| Aug 26 | 10 | 0 | 0 | 0 |
| Sep 26 | 16 | 0 | 0 | 0 |
| Oct 26 | 10 | 0 | 0 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
- Dentsply286
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Poor Quality Image205
- Communication or Transmission Problem59
- Failure to Transmit Record50
7 more in this ranking with a free account
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions338
- Insufficient Information4
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-1020-2013 · Started Sep 26, 2012 · Posted by FDA Apr 17, 2013
Reason given: It was discovered at the WEAC Labs method sample did not meet requirements for labeling and certification
Cause as FDA categorized it: Radiation Control for Health and Safety Act
Z-0999-2013 · Started Sep 18, 2012 · Posted by FDA May 2, 2013
Reason given: There have been approximately 2% of the x-ray generators manufactured between February 2012 and November 2012 returned due to errors. These generators showed the type of defect that indicates poor contact of HT transformer power pin. The defect is a bent connector associated with the generator.
Cause as FDA categorized it: Radiation Control for Health and Safety Act
Z-1000-2013 · Started Sep 18, 2012 · Posted by FDA May 2, 2013
Reason given: There have been approximately 2% of the x-ray generators manufactured between February 2012 and November 2012 returned due to errors. These generators showed the type of defect that indicates poor contact of HT transformer power pin. The defect is a bent connector associated with the generator.
Cause as FDA categorized it: Radiation Control for Health and Safety Act
3 more recall records with a free account
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.