Device type · BZO
Set, Tubing And Support, Ventilator (W Harness)
Last checked: Nov 20, 2026Source: openFDAMethods & limitations
Product code
BZO
Anesthesiology
Device class
Class I
FDA decisions since 2015
0
510(k), De Novo and PMA
Adverse event reports
175
Received Nov 1, 2025 to Nov 20, 2026
How FDA classifies it
- Device type
- Set, Tubing And Support, Ventilator (W Harness)
- Product code
- BZO
- Device class
- Class I
- Review panel
- Anesthesiology
- Regulation
- 21 CFR 868.5975
- Specialty
- Anesthesiology & Respiratory, Pulmonology
- FDA record
- Classification database entry for BZO
Latest FDA decisions
No 510(k), De Novo or PMA decisions for this product code since 2015.
Reports received
175
Received Nov 1, 2025 to Nov 20, 2026. Includes 2 of another or unstated type.
Malfunction reports
167
95% of reports
Injury reports
6
3.4% of reports
Death reports
0
0.0% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 3 | 0 | 0 | 0 |
| Jan 24 | 7 | 0 | 0 | 0 |
| Feb 24 | 2 | 0 | 0 | 0 |
| Mar 24 | 2 | 0 | 0 | 0 |
| Apr 24 | 0 | 0 | 0 | 0 |
| May 24 | 2 | 0 | 0 | 0 |
| Jun 24 | 0 | 0 | 0 | 0 |
| Jul 24 | 1 | 0 | 0 | 0 |
| Aug 24 | 4 | 0 | 0 | 0 |
| Sep 24 | 1 | 1 | 0 | 0 |
| Oct 24 | 10 | 0 | 0 | 0 |
| Nov 24 | 5 | 3 | 0 | 0 |
| Dec 24 | 7 | 0 | 0 | 0 |
| Jan 25 | 10 | 0 | 0 | 0 |
| Feb 25 | 4 | 0 | 0 | 0 |
| Mar 25 | 7 | 0 | 0 | 0 |
| Apr 25 | 10 | 0 | 0 | 0 |
| May 25 | 0 | 0 | 0 | 0 |
| Jun 25 | 8 | 0 | 0 | 0 |
| Jul 25 | 3 | 2 | 0 | 0 |
| Aug 25 | 2 | 1 | 0 | 0 |
| Sep 25 | 5 | 0 | 0 | 0 |
| Oct 25 | 7 | 1 | 0 | 0 |
| Nov 25 | 5 | 0 | 0 | 0 |
| Dec 25 | 5 | 1 | 0 | 0 |
| Jan 26 | 3 | 0 | 0 | 0 |
| Feb 26 | 5 | 0 | 0 | 0 |
| Mar 26 | 16 | 1 | 0 | 0 |
| Apr 26 | 18 | 0 | 0 | 0 |
| May 26 | 55 | 1 | 0 | 0 |
| Jun 26 | 25 | 0 | 0 | 0 |
| Jul 26 | 9 | 0 | 0 | 0 |
| Aug 26 | 9 | 2 | 0 | 0 |
| Sep 26 | 13 | 0 | 0 | 2 |
| Oct 26 | 4 | 1 | 0 | 0 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
5 more in this ranking with a free account
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- Obstruction of Flow101
- Material Split, Cut or Torn35
- Physical Resistance/Sticking24
7 more in this ranking with a free account
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions162
- Low Oxygen Saturation8
- Airway Obstruction3
3 more in this ranking with a free account
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-2992-2026 · Started Jul 13, 2026 · Posted by FDA Aug 21, 2026
Reason given: Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions,...
Cause as FDA categorized it: Under Investigation by firm
- Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.Open, Classified
Z-2463-2026 · Started May 29, 2026 · Posted by FDA Jun 30, 2026
Reason given: Ventilator coaxial breathing circuit may have, expiratory valve sets with membranes sticking to valve body, not detectable during pre-operative test, which will result in "exhalation obstructed " alarm, which may result in inadequate ventilation, oxygen desaturation. Advising: Prior to use, test lung action. If alarm...
Cause as FDA categorized it: Component design/selection
Z-0031-2026 · Started Sep 17, 2025 · Posted by FDA Oct 14, 2025
Reason given: IFU update; in-line valve, with pressure relief seal may leak when used with ventilator; don't use in neonates/infants, with target tidal volumes less-than50mL or less-than10kg- at high risk of cardiopulmonary/neurological compromise due to unrecognized hypoventilation, updates for continued use in others before...
Cause as FDA categorized it: Device Design
20 more recall records with a free account
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.