Device type · BTI
Compressor, Air, Portable
Last checked: Nov 19, 2026Source: openFDAMethods & limitations
Product code
BTI
Anesthesiology
Device class
Class II
FDA decisions since 2015
0
510(k), De Novo and PMA
Adverse event reports
3,421
Received Nov 1, 2025 to Nov 19, 2026
How FDA classifies it
- Device type
- Compressor, Air, Portable
- Product code
- BTI
- Device class
- Class II
- Review panel
- Anesthesiology
- Regulation
- 21 CFR 868.6250
- Specialty
- Anesthesiology & Respiratory
- FDA record
- Classification database entry for BTI
Latest FDA decisions
No 510(k), De Novo or PMA decisions for this product code since 2015.
Reports received
3,421
Received Nov 1, 2025 to Nov 19, 2026. Includes 17 of another or unstated type.
Malfunction reports
3,376
99% of reports
Injury reports
27
0.8% of reports
Death reports
1
0.0% of reports
Adverse event reports received per month
By the date FDA received each report, stacked by event type. The newest month is incomplete.
- Malfunction
- Injury
- Death
- Other
View as table
| Month | Malfunction | Injury | Death | Other |
|---|---|---|---|---|
| Dec 23 | 26 | 2 | 0 | 0 |
| Jan 24 | 20 | 1 | 0 | 0 |
| Feb 24 | 32 | 1 | 0 | 0 |
| Mar 24 | 3 | 3 | 0 | 0 |
| Apr 24 | 5 | 4 | 0 | 0 |
| May 24 | 4 | 2 | 0 | 0 |
| Jun 24 | 4 | 3 | 8 | 0 |
| Jul 24 | 6 | 3 | 1 | 0 |
| Aug 24 | 8 | 4 | 0 | 0 |
| Sep 24 | 2 | 3 | 8 | 0 |
| Oct 24 | 5 | 6 | 0 | 0 |
| Nov 24 | 5 | 4 | 1 | 0 |
| Dec 24 | 13 | 6 | 1 | 0 |
| Jan 25 | 2 | 7 | 1 | 0 |
| Feb 25 | 8 | 7 | 2 | 0 |
| Mar 25 | 2 | 3 | 1 | 0 |
| Apr 25 | 2 | 5 | 0 | 0 |
| May 25 | 8 | 4 | 0 | 0 |
| Jun 25 | 5 | 3 | 1 | 0 |
| Jul 25 | 5 | 0 | 0 | 0 |
| Aug 25 | 5 | 4 | 0 | 0 |
| Sep 25 | 7 | 3 | 0 | 0 |
| Oct 25 | 4 | 9 | 0 | 0 |
| Nov 25 | 5 | 7 | 0 | 0 |
| Dec 25 | 8 | 2 | 0 | 0 |
| Jan 26 | 10 | 3 | 0 | 0 |
| Feb 26 | 13 | 3 | 0 | 0 |
| Mar 26 | 24 | 3 | 0 | 0 |
| Apr 26 | 33 | 2 | 0 | 0 |
| May 26 | 92 | 0 | 0 | 0 |
| Jun 26 | 62 | 0 | 0 | 0 |
| Jul 26 | 77 | 1 | 0 | 1 |
| Aug 26 | 2,131 | 5 | 0 | 9 |
| Sep 26 | 610 | 1 | 1 | 6 |
| Oct 26 | 311 | 0 | 0 | 1 |
| Nov 26 | 0 | 0 | 0 | 0 |
Manufacturers named most often
Last 12 months. Bigger sellers get more reports. This is not a ranking of safety.
- Medline Industries3,382
- Unk1
Device problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Flow656
- Restricted Flow rate522
- Infusion or Flow Problem384
- Pressure Problem371
- Decrease in Pressure309
- Noise, Audible284
- Failure to Power Up185
- Gas Output Problem153
- Break150
- Detachment of Device or Device Component140
Patient problems as coded
Last 12 months. FDA's terms. A report can carry several.
- No Clinical Signs, Symptoms or Conditions3,341
- Insufficient Information34
- Dyspnea13
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available6
- Chronic Obstructive Pulmonary Disease (COPD)6
- Unspecified Respiratory Problem6
- Asthma5
- Cough5
- Bronchitis2
- Cancer2
Recent recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
- EVair air compressor, model numbers: M1230849 and M1230847Open, Classified
Z-0774-2024 · Started Dec 29, 2023 · Posted by FDA Jan 30, 2024
Reason given: GE HealthCare has become aware of elevated levels of formaldehyde when the EVair or EVair 03 (Jun-Air) optional compressors are used with the CARESCAPE R860 or Engstrm Carestation/Pro ventilators, respectively. These elevated results were observed in preliminary testing that was conducted at an elevated room...
Cause as FDA categorized it: Component design/selection
Z-0775-2024 · Started Dec 29, 2023 · Posted by FDA Jan 30, 2024
Reason given: GE HealthCare has become aware of elevated levels of formaldehyde when the EVair or EVair 03 (Jun-Air) optional compressors are used with the CARESCAPE R860 or Engstrm Carestation/Pro ventilators, respectively. These elevated results were observed in preliminary testing that was conducted at an elevated room...
Cause as FDA categorized it: Component design/selection
Z-0776-2024 · Started Dec 29, 2023 · Posted by FDA Jan 30, 2024
Reason given: GE HealthCare has become aware of elevated levels of formaldehyde when the EVair or EVair 03 (Jun-Air) optional compressors are used with the CARESCAPE R860 or Engstrm Carestation/Pro ventilators, respectively. These elevated results were observed in preliminary testing that was conducted at an elevated room...
Cause as FDA categorized it: Component design/selection
Maquet Cardiovascular Us Sales, Llc
Z-0848-2014 · Started Dec 5, 2013 · Posted by FDA Jan 28, 2014
Reason given: Maquet Compressor Mini may deliver compressed air at a temperature higher than specified.
Cause as FDA categorized it: Material/Component Contamination
- Medical Air Compressor Model #8413419Terminated
Z-0674-2007 · Started Jan 26, 2007 · Posted by FDA Apr 5, 2007
Reason given: Compressors not providing sufficient supply pressure to connected medical ventilator
Cause as FDA categorized it: Other
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.