FDA Clearance Tracker · P960058-S167
HiResolution Bionic Ear System
Last checked: Nov 20, 2026Latest source record: Jul 22, 2026 (this decision)Source: openFDAMethods & limitations
Decided
Jul 22, 2026
Approved
Time to decision
539 days
Received Jan 29, 2025. Calendar days, including time FDA spent waiting on the applicant.
Pathway
PMA supplement
Panel Track
Device class
Class III
The decision
- FDA number
- P960058-S167
- Device name
- HiResolution Bionic Ear System
- Applicant
- Advanced Bionics
- Applicant location
- Valencia, CA
- Received
- Jan 29, 2025
- Decided
- Jul 22, 2026
- Decision
- Approved
- Supplement type
- Panel Track
- Reason for supplement
- Labeling Change - Indications/instructions/shelf life/tradename
- Original approval
- P960058
Device type
- Device type
- Implant, cochlear (MCM)
- Device class
- Class III
- Specialty
- Otolaryngology (ENT)
FDA's approval statement
As written in FDA's record.
Unilateral Hearing Loss The HiResolution Bionic Ear System is indicated for individuals with unilateral hearing loss who meet the following criteria: Asymmetric Hearing Loss (AHL) Individuals with AHL are defined as those with a mild to moderately severe hearing loss in the better ear, i.e., a 4PTA of 31 to 65 dB HL at 500 Hz, 1000 Hz, 2000 Hz and 4000 Hz, with a difference of at least 15 dB in the 4PTA between ears. It is recommended that individuals with AHL have at least two (2) weeks experience wearing an appropriately fitted Contralateral Routing of Signal (CROS) system or suitable hearing device prior to cochlear implantation. The ear to be implanted will be as follows: Adults (18 years of age or older): Severe to profound sensorineural hearing loss (4PTA of 500, 1000, 2000, and 4000 Hz = 80 dB HL) A score of = 20% on a Consonant-Nucleus-Consonant (CNC) word test Children (5 years through 17 years of age): Profound sensorineural hearing loss (4PTA of 500, 1000, 2000, and 4000 Hz = 90 dB HL) Insufficient functional access to sound in the ear to be implanted must be determined by aided speech perception test scores of 5% or less on a developmentally appropriate monosyllabic word list when tested in the ear to be implanted alone. Single-Sided Deafness (SSD) Individuals with SSD are defined as those with normal hearing or mild sensorineural hearing loss in the better ear (4PTA of = 30 dB HL at 500, 1000, 2000, and 4000 Hz). It is recommended that individuals with SSD have at least two (2) weeks experience wearing an appropriately fitted Contralateral Routing of Signal (CROS) system or suitable hearing device prior to cochlear implantation. The ear to be implanted will be as follows: Adults (18 years of age or older): Severe to profound sensorineural hearing loss (4PTA of 500, 1000, 2000, and 4000 Hz = 80 dB HL) A score of = 20% on a Consonant-Nucleus-Consonant (CNC) word test Children (5 years through 17 years of age): Profound sensorineural hearing loss (4PTA of 500, 1000, 2000, and 4000 Hz = 90 dB HL) Insufficient functional access to sound in the ear to be implanted must be determined by aided speech perception test scores of 5% or less on a developmentally appropriate monosyllabic word list when tested in the ear to be implanted alone.
Source documents
Other decisions for Advanced Bionics
Matched on the applicant name as FDA recorded it. A company that files under several names will have more.
No other decisions since 2015 under this exact applicant name.
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.