SYNCHROMED II PROGRAMMABLE INFUSION PUMP
Last checked: Nov 20, 2026Latest source record: Nov 24, 2015 (this decision)Source: openFDAMethods & limitations
Decided
Nov 24, 2015
Approved
Time to decision
267 days
Received Mar 2, 2015. Calendar days, including time FDA spent waiting on the applicant.
Pathway
PMA supplement
Normal 180 Day Track
Device class
Class III
The decision
- FDA number
- P860004-S224
- Device name
- SYNCHROMED II PROGRAMMABLE INFUSION PUMP
- Applicant
- Medtronic, Inc.
- Applicant location
- Minneapolis, MN
- Received
- Mar 2, 2015
- Decided
- Nov 24, 2015
- Decision
- Approved
- Supplement type
- Normal 180 Day Track
- Reason for supplement
- Change Design/Components/Specifications/Material
- Original approval
- P860004
Device type
- Device class
- Class III
- Specialty
- Hospital Medicine & Nursing
FDA's approval statement
As written in FDA's record.
APPROVAL FOR TWO DESIGN CHANGES: 1) ENCAPSULATION OF MOTOR FEEDTHROUGHS WITH MEDICAL ADHESIVE IN ORDER TO REDUCE SYSTEM FAILURES FROM FEEDTHROUGH SHORTING; AND 2) THE ADDITION OF AN IMPEDANCE REQUIREMENT TO ENSURE ELECTRICAL ISOLATION OF THE MOTOR CIRCUIT.
Source documents
Other decisions for Medtronic, Inc.
Matched on the applicant name as FDA recorded it. A company that files under several names will have more.
K263596 · Decided Oct 29, 2026 · 29 days after receipt
Device type: Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass (DWF) · Cardiology
K262913 · Decided Sep 17, 2026 · 29 days after receipt
Device type: Stents, Drains And Dilators For The Biliary Ducts (FGE) · GI, Urology & Nephrology
K262735 · Decided Aug 28, 2026 · 28 days after receipt
Device type: Recorder, Event, Implantable Cardiac, (With Arrhythmia Detection) (MXD) · Cardiology
5 more FDA decisions with a free account
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.