All In MedTech

FDA Clearance Tracker · P850007-S027

PHYSIO-STIM MODELS 3202, 3302, 3313, 3314, 3315 AND SPINAL-STIM LODEL 2212

A PMA supplement is a change to a device that already has premarket approval, such as a new indication, a design change or a manufacturing change.

Last checked: Nov 20, 2026Latest source record: Apr 5, 2005 (this decision)Source: openFDAMethods & limitations

PMA supplement

Decided

Apr 5, 2005

Approved

Time to decision

671 days

Received Jun 4, 2003. Calendar days, including time FDA spent waiting on the applicant.

Pathway

PMA supplement

Normal 180 Day Track

Device class

Class II

The decision

FDA number
P850007-S027
Device name
PHYSIO-STIM MODELS 3202, 3302, 3313, 3314, 3315 AND SPINAL-STIM LODEL 2212
Applicant location
Lewisville, TX
Received
Jun 4, 2003
Decided
Apr 5, 2005
Decision
Approved
Supplement type
Normal 180 Day Track
Reason for supplement
Change Design/Components/Specifications/Material
Original approval
P850007

Device type

Device class
Class II
Regulation
21 CFR 890.5870

Every decision under product code LOF

FDA's approval statement

As written in FDA's record.

APPROVAL TO MODIFY THE PHYSIO-STIM LITE, MODELS 202L, 303L, 215L, 313L, 314L AND SPINAL-STIM MODELS 212L AND 212LE AS FOLLOWS: 1) CHANGE THE POWER SOURCE FROM A 9-VOLT DISPOSABLE BATTERY TO AN 11.1-VOLT RECHARGEABLE; 2) CHANGE THE USER INTERFACE FROM A THREE COLOR LED TO A BACKLIT LCD DISPLAY; 3) CHANGE THE CIRCUIT ASSEMBLY FROM SEPARATE BOARDS FOR THE DRIVE CIRCUIT AND MICROCONTROLLER TO A SINGLE BOARD; 4) ADD AN OPTIONAL PERSONAL DATA ASSISTANT (PDA) THAT MAY BE USED TO SET DAILY SHUTDOWN TIME FOR ORTHOFIX PERSONNEL USE ONLY; AND 5) CHANGE THE SERIAL PORT CABLE ON THE COMPLIANCE PRINTER TO AN INFRARED PORT. THE DEVICES, AS MODIFIED, WILL BE MARKETED UNDER THE TRADE NAMES PHYSIO-STIM MODELS 3202, 3302, 3313, 3314, 3315 ANDSPINAL-STIM MODEL 2212. THE PHYSIO-STIM MODELS ARE INDICATED FOR THE TREATMENT OF AN ESTABLISHED NONUNION ACQUIRED SECONDARY TO TRAUMA, EXCLUDING VERTEBRAE AND ALL FLAT BONES, WHERE THE WIDTH OF THE NONUNION DEFECTS IS LESS THAN ONE-HALF THE WIDTH OF THE BONE TO BE TREATED. A NONUNION IS CONSIDERED TO BE ESTABLISHED WHEN THE FRACTURE SITE SHOWS NO VISIBLY PROGRESSIVE SIGNS OF HEALING. THE SPINAL-STIM IS A NONINVASIVE ELECTROMAGNETIC BONE GROWTH STIMULATOR INDICATED AS A SPINAL FUSION ADJUNCT TO INCREASE THE PROBABILITY OF FUSION SUCCESS AND AS A NONOPERATIVE TREATMENT FOR SALVAGE OF FAILED SPINAL FUSION, WHERE A MINIMUM OF NINE MONTHS HAS ELAPSED SINCE LAST SURGERY.

Source documents

Other decisions for Orthofix, Inc.

Matched on the applicant name as FDA recorded it. A company that files under several names will have more.

Company page

7 more FDA decisions with a free account

For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.