FDA Clearance Tracker · P200010
Guardant360 CDx
Last checked: Nov 20, 2026Latest source record: Aug 7, 2020 (this decision)Source: openFDAMethods & limitations
Decided
Aug 7, 2020
Approved
Time to decision
179 days
Received Feb 10, 2020. Calendar days, including time FDA spent waiting on the applicant.
Pathway
PMA
Device class
Class III
The decision
- FDA number
- P200010
- Device name
- Guardant360 CDx
- Applicant
- Guardant Health, Inc.
- Applicant location
- Redwood City, CA
- Received
- Feb 10, 2020
- Decided
- Aug 7, 2020
- Decision
- Approved
- Supplements
- 30 in FDA's data for this PMA. See them on the FDA record
Device type
- Device type
- Next generation sequencing oncology panel, somatic or germline variant detection system (PQP)
- Device class
- Class III
- Specialty
- Laboratory Medicine & Pathology
FDA's approval statement
As written in FDA's record.
Approval order for Guardant360® CDx. Guardant360® CDx is a qualitative next generation sequencing-based in vitro diagnostic device that uses targeted high throughput hybridization-based capture technology for detection of single nucleotide variants (SNVs), insertions and deletions (indels) in 55 genes, copy number amplifications (CNAs) in two (2) genes, and fusions in four (4) genes. Guardant360 CDx utilizes circulating cell-free DNA (cfDNA) from plasma of peripheral whole blood collected in Streck Cell-Free DNA Blood Collection Tubes (BCTs). The test is intended to be used as a companion diagnostic to identify non-small cell lung cancer (NSCLC) patients who may benefit from treatment with the targeted therapy listed in Table 1 in accordance with the approved therapeutic product labeling. Table 1. Companion Diagnostic IndicationsIndication: Non-small cell lung cancer (NSCLC); Biomarker: EGFR exon 19 deletions, L858R, and T790M*; Therapy: TAGRISSO® (osimertinib)A negative result from a plasma specimen does not assure that the patients tumor is negative for genomic findings. NSCLC patients who are negative for the biomarkers listed in Table 1 should be reflexed to tissue biopsy testing for Table 1 biomarkers using an FDA-approved tumor tissue test, if feasible.*The efficacy of TAGRISSO® (osimertinib) has not been established in the EGFR T790M plasma-positive, tissue-negative or unknown population and clinical data for T790M plasma-positive patients are limited; therefore, testing using plasma specimens is most appropriate for consideration in patients from whom a tumor biopsy cannot be obtained.Additionally, the test is intended to provide tumor mutation profiling to be used by qualified health care professionals in accordance with professional guidelines in oncology for cancer patients with solid malignant neoplasms. The test is for use with patients previously diagnosed with cancer and in conjunction with other laboratory and clinical findings. Genomic findings other than those listed in Table 1 are not prescriptive or conclusive for labeled use of any specific therapeutic product. Guardant360 CDx is a single-site assay performed at Guardant Health, Inc.
How new is this?
These are facts about FDA's records. They are not a judgment about the device.
- It is not the first decision under product code PQP. The earliest in FDA's data is P160018, decided Dec 19, 2016.
- It is on FDA's public list of Breakthrough Devices with marketing authorization (list dated June 30, 2026).
Source documents
Other decisions for Guardant Health, Inc.
Matched on the applicant name as FDA recorded it. A company that files under several names will have more.
P250027 · Decided May 19, 2026 · 292 days after receipt
Device type: Next generation sequencing oncology panel, somatic or germline variant detection system (PQP) · Laboratory Medicine & Pathology
- Guardant360 CDxPMA supplement
P200010-S026 · Decided Jan 15, 2026 · 176 days after receipt · Panel Track
Device type: Next generation sequencing oncology panel, somatic or germline variant detection system (PQP) · Laboratory Medicine & Pathology
- Guardant360 CDxPMA supplement
P200010-S021 · Decided Sep 25, 2025 · 325 days after receipt · Panel Track
Device type: Next generation sequencing oncology panel, somatic or germline variant detection system (PQP) · Laboratory Medicine & Pathology
5 more FDA decisions with a free account
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.