FDA Clearance Tracker · P060019-S013
THERAPY COOL PATH ABLATION CATHETER
Last checked: Nov 20, 2026Latest source record: Dec 3, 2010 (this decision)Source: openFDAMethods & limitations
Decided
Dec 3, 2010
Approved
Time to decision
95 days
Received Aug 30, 2010. Calendar days, including time FDA spent waiting on the applicant.
Pathway
PMA supplement
Normal 180 Day Track
Device class
Class III
The decision
- FDA number
- P060019-S013
- Device name
- THERAPY COOL PATH ABLATION CATHETER
- Applicant
- Irvine Biomedical, Inc.
- Applicant location
- Irvine, CA
- Received
- Aug 30, 2010
- Decided
- Dec 3, 2010
- Decision
- Approved
- Supplement type
- Normal 180 Day Track
- Reason for supplement
- Change Design/Components/Specifications/Material
- Original approval
- P060019
Device type
- Device class
- Class III
- Specialty
- CardiologyElectrophysiology
FDA's approval statement
As written in FDA's record.
APPROVAL TO CHANGE THE MATERIAL OF THE EXTENSION TUBE TO INCREASE THE TORQUE RESISTANCE OF THE EXTENSION TUBE DURING HANDLE MANIPULATION. THE MATERIAL OF THE EXTENSION TUBE WILL BE CHANGED FROM A NON-BRAIDED PELLETHANE TO A PEBAX BRAIDED TUBE (72D). ADDITIONALLY, THE LENGTH OF THE EXTENSION TUBE AND THE INTERNAL FLUID LUMEN TUBE WILL BE DECREASED FROM 11 CM TO 4 CM.
Source documents
Other decisions for Irvine Biomedical, Inc.
Matched on the applicant name as FDA recorded it. A company that files under several names will have more.
No other decisions since 2015 under this exact applicant name.
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.