PMA device
Pipeline Embolization Devices,pipeline Flex Embolization Device
Micro Therapeutics, Inc. d/b/a ev3 Neurovascular · Intracranial Aneurysm Flow Diverter (OUT, Class III) · First FDA decision Apr 6, 2011, newest Jan 26, 2015Brain aneurysm
Last checked: Nov 20, 2026Source: FDA, CMS, ClinicalTrials.gov, PubMed
FDA decisions
4
PMA pathway
Device trials
18
Naming "Pipeline Embolization"
MAUDE reports
672
Brand "Pipeline Embolization", since 2015
PubMed studies
1,027
Mentioning "Pipeline Embolization"
AI summary across all sources
Written by AI from the FDA, CMS, ClinicalTrials.gov, MAUDE and PubMed data on this page, plus web searches for payment facts.
Product review deck
A slide deck reviewing Pipeline Embolization Devices,pipeline Flex Embolization Device
About ten slides on the technology, clinical evidence, regulatory history and commercialization, built with AI from the data on this page and web research, with sources. Download it as PowerPoint to edit or as PDF to share.
- Executive summary and key facts
- Technology
- Clinical evidence and trials
- Regulatory history and safety
- Reimbursement, market and competition
- Risks, outlook and sources
AI-written from public data and may contain mistakes. Check each fact against its source before relying on it. Not investment, legal, reimbursement or medical advice.
FDA decisions
The newest 4 of 4 decisions filed under this name.
P100018-S011 · Decided Jan 26, 2015 · 214 days after receipt · Normal 180 Day Track
Device type: Intracranial aneurysm flow diverter (OUT) · Neurology & Neurosurgery
- PIPELINE EMBOLIZATION DEVICEPMA supplement
P100018-S004 · Decided Mar 22, 2012 · 185 days after receipt · Normal 180 Day Track No User Fee
Device type: Intracranial aneurysm flow diverter (OUT) · Neurology & Neurosurgery
- PIPELINE EMBOLIZATION DEVICEPMA supplement
P100018-S001 · Decided Jun 27, 2011 · 60 days after receipt · Normal 180 Day Track No User Fee
Device type: Intracranial aneurysm flow diverter (OUT) · Neurology & Neurosurgery
1 more FDA decisions with a free account
Medicare coverage
National and local coverage documents whose titles match this device or its device type. Payment add-ons (NTAP, pass-through, TCET) and likely codes are covered in the summary above.
No Medicare coverage document has a title matching this device or device type. Many devices are paid under existing codes without one.
Clinical trials
Device studies on ClinicalTrials.gov that name "Pipeline Embolization", any status and sponsor, newest update first.
Medtronic Neurovascular Clinical AffairsPublic (MDT)
NCT04391803 · 25 participants · Start May 6, 2022 · Primary completion Oct 20, 2025
Device: Pipeline™ Flex Embolization Device with Shield Technology™ · For: Intracranial Aneurysm; Ruptured Aneurysm
Sites: New Haven, Connecticut; Marietta, Georgia; Chicago, Illinois · 7 sites in all
Medtronic Neurovascular Clinical AffairsPublic (MDT)
NCT06604884 · 118 participants (planned) · Start Dec 12, 2025 · Primary completion Aug 2029
Device: Treatment of Intracranial Aneurysms · For: Intracranial Aneurysm
Sites: Jacksonville, Florida; Marietta, Georgia; Evanston, Illinois · 10 sites in all
- INSPIRE Pipeline™ Shield Post Approval StudyActive, not recruiting
Medtronic Neurovascular Clinical AffairsPublic (MDT)
NCT05071963 · 246 participants · Start Dec 20, 2021 · Primary completion Jan 2027
Device: Treatment of Intracranial Aneurysms · For: Intracranial Aneurysm
Sites: Birmingham, Alabama; New Haven, Connecticut; Jacksonville, Florida · 16 sites in all
15 more trials with a free account
Adverse event reports (MAUDE)
Reports received since 2015 naming the brand "Pipeline Embolization", and for the whole product code for context.
This brand
672 reports: 161 malfunction, 438 injury, 72 death
Whole product code OUT
10,961 reports: 7,308 malfunction, 3,152 injury, 499 death
Newest reports for this brand
- PIPELINE EMBOLIZATION DEVICE (PED)Malfunction
12405433 · Received Sep 1, 2021 · Event dated Aug 10, 2021
Device type: Intracranial Aneurysm Flow Diverter (OUT)
Device problem as coded: Activation, Positioning or Separation Problem
Patient problem as coded: No Clinical Signs, Symptoms or Conditions
EV3 (NOW COVIDIEN-MEDTRONICS) / MEDTRONIC INC, -SWINNEA
MW5099621 · Received Feb 24, 2021 · Event dated Apr 4, 2021
Device type: Intracranial Aneurysm Flow Diverter (OUT)
Device problem as coded: Insufficient Device Problem Information
Patient problem as coded: Aneurysm; Stroke/CVA; Headache; Occlusion; Implant Pain
MICRO THERAPEUTICS, INC. DBA EV3
2029214-2020-00240 · Received Mar 17, 2020 · Event dated Apr 15, 2015
Device type: Intracranial Aneurysm Flow Diverter (OUT)
Device problem as coded: No Apparent Adverse Event
Patient problem as coded: Death
2 more reports with a free account
Recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-1245-2025 · Started Jan 30, 2025 · Posted by FDA Mar 5, 2025
Reason given: Use of embolization device with diameters greater-than-or-equal-to 4mm may lead to incomplete wall apposition and/or braid deformation (fish-mouthing/braid narrowing/braid collapse) noted during procedure and post-procedure, typically noted at 6-12 month imaging follow-up, potentially can lead to thrombosis and/or...
Cause as FDA categorized it: Device Design · Device type: Intracranial Aneurysm Flow Diverter (OUT)
Z-1246-2025 · Started Jan 30, 2025 · Posted by FDA Mar 5, 2025
Reason given: Use of embolization device with diameters greater-than-or-equal-to 4mm may lead to incomplete wall apposition and/or braid deformation (fish-mouthing/braid narrowing/braid collapse) noted during procedure and post-procedure, typically noted at 6-12 month imaging follow-up, potentially can lead to thrombosis and/or...
Cause as FDA categorized it: Device Design · Device type: Intracranial Aneurysm Flow Diverter (OUT)
- Pipeline Flex Embolization DeviceOpen, Classified
Z-2402-2021 · Started Jul 13, 2021
Reason given: Due to potential push wire fractures in the delivery system during use. The issue presents as a fracture in the spiral cut area of the Hypotube.
Cause as FDA categorized it: Device Design · Device type: Intracranial Aneurysm Flow Diverter (OUT)
9 more recall records with a free account
Published research
The most relevant PubMed records for "Pipeline Embolization"[tiab]. The summary above synthesizes these.
- Pipeline embolization of proximal middle cerebral artery aneurysms: A multicenter cohort study.
Lauzier DC, Root BK, Kayan Y, et al. · Interv Neuroradiol · 2022Multicenter Study
- Pipeline embolization of complex, wide-necked middle cerebral artery bifurcation aneurysms: A single-center experience.
Wang J, Deng X, Li D, et al. · Interv Neuroradiol · 2024
- Pipeline embolization devices for the treatment of nonsaccular aneurysms in pediatric patients.
Han J, Liang F, Zhang Y, et al. · Front Neurol · 2023
- Pipeline embolization device dynamics: prediction of incomplete occlusion by elongation from nominal length.
Sunohara T, Imamura H, Ohta T, et al. · J Neurosurg · 2024
- Clinical Management of Cerebral Aneurysms-Endoluminal.
Shapiro M, Nossek E, Sharashidze V, et al. · Neuroimaging Clin N Am · 2025Review
- Pipeline embolization in patients with hemoglobinopathies: A cohort study.
Grin EA, Sharashidze V, Shapiro M, et al. · Interv Neuroradiol · 2025
- Flow-diversion panacea or poison?
Zanaty M, Chalouhi N, Tjoumakaris SI, et al. · Front Neurol · 2014Review
- Pipeline embolization of MCA aneurysms in the M2-M4 segment: Dual center study and meta-analysis.
Lauzier DC, Root BK, Kayan Y, et al. · Clin Neurol Neurosurg · 2022Meta-Analysis, Multicenter Study
- Intra-aneurysmal thrombus modification after flow-diversion.
Zanaty M, Jabbour PM, Bou Sader R, et al. · J Clin Neurosci · 2015
- Pipeline Embolization Device for intracranial aneurysms presenting with mass effect: a large Chinese cohort.
Zhao Y, Lu J, Zhang H, et al. · Stroke Vasc Neurol · 2024Research Support, Non-U.S. Gov't
- Clinical outcomes of pipeline embolization devices with shield technology for treating intracranial aneurysms.
Luo C, Jin L, Dong J, et al. · Front Neurol · 2022Systematic Review
- Initial experience with Pipeline embolization of intracranial pseudoaneurysms in pediatric patients.
Budohoski KP, Thakrar R, Voronovich Z, et al. · J Neurosurg Pediatr · 2022
Other devices of this type
Also filed under product code OUT.
- Pipeline Flex Embolization Device with Shield TechnologyMicro Therapeutics, Inc. d/b/a ev3 Neurovascular · PMA 2021
- Flow Re-Direction Endoluminal Device (FRED®) SystemMicroVention, Inc. · PMA 2025
- Pipeline Vantage Embolization Device with Shield TechnologyMicro Therapeutics, Inc. d/b/a ev3 Neurovascular · PMA 2025
- Surpass Streamline Flow DiverterStryker Neurovascular · PMA 2022
- Surpass Elite Flow Diverter SystemStryker Neurovascular · PMA 2024
- FRED X SystemMicroVention, Inc. · PMA 2022
- Surpass Evolve Flow Diverter SystemStryker Neurovascular · PMA 2020
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.