PMA device
X-Stop Interspinous Spacer System
Medtronic Sofamor Danek, Inc. · Prosthesis, Spinous Process Spacer/Plate (NQO, Class III) · First FDA decision Oct 4, 2007, newest Jul 25, 2012Degenerative disc disease and spinal stenosis
Last checked: Nov 20, 2026Source: FDA, CMS, ClinicalTrials.gov, PubMed
FDA decisions
7
PMA pathway
Device trials
5
Naming "X-Stop Interspinous"
MAUDE reports
26
Brand "X-Stop Interspinous", since 2015
PubMed studies
26
Mentioning "X-Stop Interspinous"
AI summary across all sources
Written by AI from the FDA, CMS, ClinicalTrials.gov, MAUDE and PubMed data on this page, plus web searches for payment facts.
Product review deck
A slide deck reviewing X-Stop Interspinous Spacer System
About ten slides on the technology, clinical evidence, regulatory history and commercialization, built with AI from the data on this page and web research, with sources. Download it as PowerPoint to edit or as PDF to share.
- Executive summary and key facts
- Technology
- Clinical evidence and trials
- Regulatory history and safety
- Reimbursement, market and competition
- Risks, outlook and sources
AI-written from public data and may contain mistakes. Check each fact against its source before relying on it. Not investment, legal, reimbursement or medical advice.
FDA decisions
The newest 6 of 7 decisions filed under this name.
- X-STOP INTERSPINOUS SPACER SYSTEMPMA supplement
P040001-S020 · Decided Jul 25, 2012 · 147 days after receipt · Normal 180 Day Track No User Fee
Device type: Prosthesis, spinous process spacer/plate (NQO) · Orthopedic Surgery
- X-STOP INTERSPINOUS SPACER SYSTEMPMA supplement
P040001-S017 · Decided Jun 2, 2010 · 146 days after receipt · Normal 180 Day Track No User Fee
Device type: Prosthesis, spinous process spacer/plate (NQO) · Orthopedic Surgery
- X-STOP INTERSPINOUS SPACER SYSTEMPMA supplement
P040001-S016 · Decided Apr 30, 2010 · 176 days after receipt · Normal 180 Day Track
Device type: Prosthesis, spinous process spacer/plate (NQO) · Orthopedic Surgery
3 more FDA decisions with a free account
Older decisions: P040001-S007
Medicare coverage
National and local coverage documents whose titles match this device or its device type. Payment add-ons (NTAP, pass-through, TCET) and likely codes are covered in the summary above.
No Medicare coverage document has a title matching this device or device type. Many devices are paid under existing codes without one.
Clinical trials
Device studies on ClinicalTrials.gov that name "X-Stop Interspinous", any status and sponsor, newest update first.
Medtronic Spine LLCPublic (MDT)
NCT00534092 · 69 participants · Start Dec 2006
Device: X STOP® Interspinous Process Decompression System · For: Lumbar Spinal Stenosis
Sites: San Francisco, California; Scarborough, Maine; Timonium, Maryland · 5 sites in all
- Investigating Superion™ In Spinal StenosisPivotal-lookingCompleted
Boston Scientific CorporationPublic (BSX)
NCT00692276 · 391 participants · Start Jun 2008 · Primary completion Feb 2017
Device: Superion™ Interspinous Spacer; X-STOP® IPD® Device · For: Lumbar Spinal Stenosis; Intermittent Claudication
Sites: Phoenix, Arizona; Tucson, Arizona; Beverly Hills, California · 32 sites in all
- Condition of Approval StudyTerminated
Medtronic Spine LLCPublic (MDT)
NCT00517751 · 176 participants · Start Aug 2007 · Primary completion Feb 2015
Device: X-STOP PEEK · For: Lumbar Spinal Stenosis
Sites: Birmingham, Alabama; Tucson, Arizona; Campbell, California · 23 sites in all
2 more trials with a free account
Adverse event reports (MAUDE)
Reports received since 2015 naming the brand "X-Stop Interspinous", and for the whole product code for context.
This brand
26 reports: 2 malfunction, 24 injury, 0 death
Whole product code NQO
2,342 reports: 1,410 malfunction, 926 injury, 1 death
Newest reports for this brand
MEDTRONIC SOFAMOR DANEK USA, INC
1030489-2018-01246 · Received Sep 17, 2018
Device type: Prosthesis, Spinous Process Spacer/Plate (NQO)
Device problem as coded: Migration or Expulsion of Device
Patient problem as coded: Claudication
1030489-2018-00448 · Received Mar 26, 2018
Device type: Prosthesis, Spinous Process Spacer/Plate (NQO)
Device problem as coded: Adverse Event Without Identified Device or Use Problem
Patient problem as coded: Cerebrospinal Fluid Leakage; Bone Fracture(s); Hematoma; Unspecified Infection; Pain; Stenosis; Discomfort
- X-STOP INTERSPINOUS SPACER SYSTEMMalfunction
1030489-2018-00449 · Received Mar 26, 2018
Device type: Prosthesis, Spinous Process Spacer/Plate (NQO)
Device problem as coded: Migration or Expulsion of Device
Patient problem as coded: Discomfort
2 more reports with a free account
Recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-1765-2008 · Started Jan 30, 2008 · Posted by FDA Sep 29, 2008
Reason given: Physician instructions revised due to product breakage: During the implantation of the X-STOP IPD Device, the product has the potential to cause damage to and/or breakage of the X-STOP IPD Device's universal wing assembly. This includes the potential for breakage of the wing screw.
Cause as FDA categorized it: Labeling design · Device type: Prosthesis, Spinous Process Spacer/Plate (NQO)
St. Francis Medical Technologies Inc
Z-1258-06 · Started Jun 6, 2006 · Posted by FDA Jul 26, 2006
Reason given: Some units may be mislabeled as 10 mm X STOP implant that actually contain 14 mm X STOP implant.
Cause as FDA categorized it: Other · Device type: Prosthesis, Spinous Process Spacer/Plate (NQO)
Published research
The most relevant PubMed records for "X-Stop Interspinous"[tiab]. The summary above synthesizes these.
- Effectiveness of X-stop Interspinous Distractor Device Versus Laminectomy for Treatment of Lumbar Stenosis: A Systematic Review and Meta-Analysis.
Ekhator C, Griepp D, Urbi A, et al. · Cureus · 2023Review
- High failure rate of the interspinous distraction device (X-Stop) for the treatment of lumbar spinal stenosis caused by degenerative spondylolisthesis.
Verhoof OJ, Bron JL, Wapstra FH, et al. · Eur Spine J · 2008
- Dynamic interspinous process stabilization: review of complications associated with the X-Stop device.
Bowers C, Amini A, Dailey AT, et al. · Neurosurg Focus · 2010
- Scoliosis and interspinous decompression with the X-STOP: prospective minimum 1-year outcomes in lumbar spinal stenosis.
Rolfe KW, Zucherman JF, Kondrashov DG, et al. · Spine J · 2010Research Support, Non-U.S. Gov't
- One-year results of X Stop interspinous implant for the treatment of lumbar spinal stenosis.
Siddiqui M, Smith FW, Wardlaw D · Spine (Phila Pa 1976) · 2007Clinical Trial, Research Support, Non-U.S. Gov't
- Cost-effectiveness of the X-STOP® interspinous spacer for lumbar spinal stenosis.
Skidmore G, Ackerman SJ, Bergin C, et al. · Spine (Phila Pa 1976) · 2011Comparative Study, Multicenter Study
- Insertion loads of the X STOP interspinous process distraction system designed to treat neurogenic intermittent claudication.
Talwar V, Lindsey DP, Fredrick A, et al. · Eur Spine J · 2006
- X-Stop interspinous implant for the treatment of lumbar spinal stenosis. Our experience after 50 consecutive patients.
Ulivieri S, Petrini C, Oliveri G · G Chir · 2010
- Two-Year Evaluation of the X-STOP Interspinous Spacer in Different Primary Patient Populations With Neurogenic Intermittent Claudication Because of Lumbar Spinal Stenosis.
Hartjen CA, Resnick DK, Hsu KY, et al. · Clin Spine Surg · 2016Multicenter Study
- Evaluation of interspinous process distraction device (X-STOP) in a representative patient cohort.
Patil S, Burton M, Storey C, et al. · World Neurosurg · 2013
- Interspinous process decompression with the X-STOP device for lumbar spinal stenosis: a 4-year follow-up study.
Kondrashov DG, Hannibal M, Hsu KY, et al. · J Spinal Disord Tech · 2006Comparative Study
- X-STOP surgical implant for the treatment of lumbar spinal stenosis: clinical practice recommendations for neurosurgical nurse practitioners.
Nielsen M · J Neurosci Nurs · 2013Review
Other devices of this type
Also filed under product code NQO.
- Superion Indirect Decompression SystemBoston Scientific Neuromodulation · PMA 2024
- coflex® Interlaminar TechnologyCompanion Spine France · PMA 2026
- Superion Interspinous SpacerBoston Scientific Neuromodulation · PMA 2017
- coflex Interlaminar Stabilization DeviceCompanion Spine France · PMA 2021
- X Stop Interspinous Process Decompression SystemMedtronic Sofamor Danek, Inc. · PMA 2006
- DIAM Spinal Stabilization SystemCompanion Spine France Sas · PMA 2025
- Paradigm Spine Coflex Interlaminar Stabilization DeviceCompanion Spine France · PMA 2018
- X-Stop LtosMedtronic Sofamor Danek, Inc. · PMA 2014
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.