PMA device
Percepta Quad CRT-P MRI SureScan
Medtronic, Inc. · Implantable Pulse Generator, Pacemaker (Non-Crt) (LWP, Class III) · First FDA decision May 6, 2017, newest Apr 10, 2019Bradycardia and heart block
Last checked: Nov 20, 2026Source: FDA, CMS, ClinicalTrials.gov, PubMed
FDA decisions
4
PMA pathway
Device trials
0
Naming "Percepta Quad CRT-P"
MAUDE reports
1,589
Brand "Percepta Quad CRT-P", since 2015
PubMed studies
0
Mentioning "Percepta Quad CRT-P"
AI summary across all sources
Written by AI from the FDA, CMS, ClinicalTrials.gov, MAUDE and PubMed data on this page, plus web searches for payment facts.
Product review deck
A slide deck reviewing Percepta Quad CRT-P MRI SureScan
About ten slides on the technology, clinical evidence, regulatory history and commercialization, built with AI from the data on this page and web research, with sources. Download it as PowerPoint to edit or as PDF to share.
- Executive summary and key facts
- Technology
- Clinical evidence and trials
- Regulatory history and safety
- Reimbursement, market and competition
- Risks, outlook and sources
AI-written from public data and may contain mistakes. Check each fact against its source before relying on it. Not investment, legal, reimbursement or medical advice.
FDA decisions
The newest 4 of 4 decisions filed under this name.
- Percepta Quad CRT-P MRI SureScan, Serena Quad CRT-P MRI SureScan, Solara Quad CRT-P MRI SureScan and Implantable Pulse GPMA supplement
P010015-S385 · Decided Apr 10, 2019 · 166 days after receipt · Normal 180 Day Track
Device type: Implantable pulse generator, pacemaker (non-CRT) (LWP) · Cardiology
- Percepta Quad CRT-P MRI SureScan, Percepta CRT-P MRI SureScan, Serena Quad CRT-P MRI SureScan, Serena CRT-P MRI SureScanPMA supplement
P010015-S368 · Decided Jun 28, 2018 · 76 days after receipt · Normal 180 Day Track
Device type: Implantable pulse generator, pacemaker (non-CRT) (LWP) · Cardiology
- Percepta Quad CRT-P MRI SureScan / Percepta CRT-P MRI SureScan / Serena Quad CRT-P MRI SureScan / Splar Quad CRT-P MRI SPMA supplement
P010015-S340 · Decided Oct 19, 2017 · 90 days after receipt · Normal 180 Day Track
Device type: Implantable pulse generator, pacemaker (non-CRT) (LWP) · Cardiology
1 more FDA decisions with a free account
Medicare coverage
National and local coverage documents whose titles match this device or its device type. Payment add-ons (NTAP, pass-through, TCET) and likely codes are covered in the summary above.
No Medicare coverage document has a title matching this device or device type. Many devices are paid under existing codes without one.
Clinical trials
Device studies on ClinicalTrials.gov that name "Percepta Quad CRT-P", any status and sponsor, newest update first.
No registered device study names "Percepta Quad CRT-P".
Adverse event reports (MAUDE)
Reports received since 2015 naming the brand "Percepta Quad CRT-P", and for the whole product code for context.
This brand
1,589 reports: 765 malfunction, 787 injury, 37 death
Whole product code LWP
142,226 reports: 58,138 malfunction, 82,585 injury, 1,486 death
Newest reports for this brand
9614453-2026-04420 · Received Oct 29, 2026 · Event dated Oct 13, 2026
Device type: Pulse Generator, Pacemaker, Implantable, With Cardiac Resynchronization (Crt-P) (NKE)
Device problem as coded: Adverse Event Without Identified Device or Use Problem
Patient problem as coded: Erythema; Unspecified Infection; Discomfort; Swelling/ Edema
- Percepta Quad CRT-P MRI SureScanMalfunction
9614453-2026-04377 · Received Oct 28, 2026 · Event dated Oct 1, 2026
Device type: Pulse Generator, Pacemaker, Implantable, With Cardiac Resynchronization (Crt-P) (NKE)
Device problem as coded: Appropriate Device Problem Term/Code Not Available
Patient problem as coded: Insufficient Information
9614453-2026-04371 · Received Oct 27, 2026 · Event dated Oct 7, 2026
Device type: Pulse Generator, Pacemaker, Implantable, With Cardiac Resynchronization (Crt-P) (NKE)
Device problem as coded: Adverse Event Without Identified Device or Use Problem
Patient problem as coded: Fever; Unspecified Infection; Sepsis; Medical device site infection; Bacteremia
2 more reports with a free account
Recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-2127-2025 · Started Jun 17, 2025 · Posted by FDA Jul 17, 2025
Reason given: A software update for Medtronic CareLink SmartSync Device Manager (SmartSync) is now available. The update eliminates the potential display of an erroneous electrical reset (pop-up) message upon interrogation with a SmartSync Device Manager. The erroneous message indicates an electrical reset has occurred. However, no...
Cause as FDA categorized it: Software design · Device type: Programmer, Pacemaker (KRG)
Z-2128-2025 · Started Jun 17, 2025 · Posted by FDA Jul 17, 2025
Reason given: A software update for Medtronic CareLink SmartSync Device Manager (SmartSync) is now available. The update eliminates the potential display of an erroneous electrical reset (pop-up) message upon interrogation with a SmartSync Device Manager. The erroneous message indicates an electrical reset has occurred. However, no...
Cause as FDA categorized it: Software design · Device type: Programmer, Pacemaker (KRG)
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Z-2511-2020 · Started Jun 4, 2020 · Posted by FDA Jul 10, 2020
Reason given: There is a rare communication sequence during the first device interrogation with a SmartSync Device Manager that may result in the temporary suspension of some device features (i.e., battery measurements, Capture Management", Atrial Lead Position Check", EffectivCRT" algorithms, and AdaptivCRT"). This rare...
Cause as FDA categorized it: Software design · Device type: Pulse Generator, Permanent, Implantable (NVZ)
7 more recall records with a free account
Published research
The most relevant PubMed records for "Percepta Quad CRT-P"[tiab]. The summary above synthesizes these.
No PubMed record mentions this device by name.
Other devices of this type
Also filed under product code LWP.
- CapSureFix MRI LeadMedtronic, Inc. · PMA 2025
- Revo MRI Surescan Ipg and Pacing SystemMedtronic, Inc. · PMA 2020
- Assurity MRI / Endurity MRI PacemakersAbbott Medical · PMA 2026
- Imageready MR Conditional Pacing System and Ingevity Pace/Sense LeadBoston Scientific · PMA 2026
- INGEVITY MRI tined and extendable/retractable fixation leads / INGEVITY+ leadsBoston Scientific · PMA 2026
- Tendril MRI Pacing Leads / Assurity MRIAbbott Medical · PMA 2025
- AdvantioBoston Scientific Corp · PMA 2018
- Acuity Spiral LeadBoston Scientific Corp · PMA 2013
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.