FDA decisions
The newest 6 of 15 decisions filed under this name.
K180906 · Decided May 16, 2018 · 40 days after receipt
Device type: Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer (JWH) · Orthopedic Surgery
K153721 · Decided Mar 25, 2016 · 88 days after receipt
Device type: Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer (JWH) · Orthopedic Surgery
K142161 · Decided Nov 14, 2014 · 100 days after receipt
Device type: Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer (JWH) · Orthopedic Surgery
K140423 · Decided May 27, 2014 · 98 days after receipt
Device type: Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer (JWH) · Orthopedic Surgery
K133560 · Decided Mar 28, 2014 · 128 days after receipt
Device type: Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer (JWH) · Orthopedic Surgery
K133256 · Decided Jan 9, 2014 · 78 days after receipt
Device type: Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer (JWH) · Orthopedic Surgery
Older decisions: K131467, K131019, K122991, K122870, K122033, K120316, K113378, K112780, K103117
Medicare coverage
National and local coverage documents whose titles match this device or its device type. Payment add-ons (NTAP, pass-through, TCET) and likely codes are covered in the summary above.
No Medicare coverage document has a title matching this device or device type. Many devices are paid under existing codes without one.
Clinical trials
Device studies on ClinicalTrials.gov that name "iTotal Cruciate Retaining", any status and sponsor, newest update first.
Restor3DNot on our public-company list
NCT01564654 · 89 participants · Start Jan 2012 · Primary completion Jan 2013
Device: iTotal KRS · For: Osteoarthritis
Sites: La Quinta, California; Boynton Beach, Florida; Nashville, Tennessee · 4 sites in all
Restor3DNot on our public-company list
NCT01899417 · 235 participants · Start Jul 2013 · Primary completion Mar 2014
Device: ConforMIS iTotal® Knee Replacement · For: Osteoarthritis,Knee
Sites: Atlantis, Florida
Restor3DNot on our public-company list
NCT03152357 · 950 participants · Start Apr 1, 2017 · Primary completion Sep 1, 2019
Device: ConforMIS iUni, iDuo, or iTotal (CR) · For: Osteoarthritis, Knee; Knee Osteoarthritis
Sites: Boynton Beach, Florida
Adverse event reports (MAUDE)
Reports received since 2015 naming the brand "iTotal Cruciate Retaining", and for the whole product code for context.
This brand
1 reports: 0 malfunction, 1 injury, 0 death
Whole product code JWH
121,905 reports: 20,134 malfunction, 101,388 injury, 234 death
Newest reports for this brand
3004153240-2024-00017 · Received Oct 18, 2024 · Event dated Oct 10, 2024
Device type: Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer (JWH)
Device problem as coded: Adverse Event Without Identified Device or Use Problem
Patient problem as coded: Unspecified Infection
Recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
Z-2849-2020 · Started Jul 20, 2020
Reason given: Outer box labels for kit 0469247 correct, however the boxes contained the implants and jigs for 0468920
Cause as FDA categorized it: Packaging process control · Device type: Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer (JWH)
Z-1627-2020 · Started Feb 3, 2020
Reason given: A limited number of tibial trays were labeled with an incorrect serial number and were delivered to the wrong customer accounts.
Cause as FDA categorized it: Mixed-up of materials/components · Device type: Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer (JWH)
Z-0672-2020 · Started Oct 25, 2019
Reason given: Medial and lateral polyethylene inserts did not securely fit into the cobalt chrome tray implant.
Cause as FDA categorized it: Under Investigation by firm · Device type: Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer (JWH)
Z-1108-2019 · Started Dec 3, 2018
Reason given: The kits were prepared with incorrect patient ID labels and may contain incorrect components.
Cause as FDA categorized it: Labeling mix-ups · Device type: Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer (JWH)
Z-1070-2016 · Started Jan 28, 2016
Reason given: Incorrect femoral component distributed
Cause as FDA categorized it: Packaging process control · Device type: Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer (JWH)
Z-0138-2016 · Started Aug 31, 2015 · Posted by FDA Oct 21, 2015
Reason given: May contain small amounts of ethylene glycol residue
Cause as FDA categorized it: Process design · Device type: Knee Arthroplasty Implantation System (OOG)
Z-0139-2016 · Started Aug 31, 2015 · Posted by FDA Oct 21, 2015
Reason given: May contain small amounts of ethylene glycol residue
Cause as FDA categorized it: Process design · Device type: Knee Arthroplasty Implantation System (OOG)
Z-0140-2016 · Started Aug 31, 2015 · Posted by FDA Oct 21, 2015
Reason given: May contain small amounts of ethylene glycol residue
Cause as FDA categorized it: Process design · Device type: Knee Arthroplasty Implantation System (OOG)
Z-0141-2016 · Started Aug 31, 2015 · Posted by FDA Oct 21, 2015
Reason given: May contain small amounts of ethylene glycol residue
Cause as FDA categorized it: Process design · Device type: Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer (JWH)
Z-0142-2016 · Started Aug 31, 2015 · Posted by FDA Oct 21, 2015
Reason given: May contain small amounts of ethylene glycol residue
Cause as FDA categorized it: Process design · Device type: Prosthesis, Knee Patellofemorotibial, Partial, Semi-Constrained, Cemented, Polymer/Metal/Polymer (NPJ)
Showing the 10 newest of 17 recall records in FDA's data.
Published research
The most relevant PubMed records for "iTotal Cruciate Retaining"[tiab]. The summary above synthesizes these.
No PubMed record mentions this device by name.
Other devices of this type
Also filed under product code JWH.
- Klassic Knee SystemTotal Joint Orthopedics, Inc. · 510(k) 2022
- Balanced Knee® System TriMax PS Plus Tibial InsertOrtho Development Corporation · 510(k) 2023
- Freedom Total Knee System (E-Poly Tibial Liners)Maxx Orthopedics, Inc. · 510(k) 2026
- Zimmer NexGen Complete Knee Solution Tibial and Femoral AugmentsZimmer, Inc. · 510(k) 2015
- Materialise TKA Guide SystemMaterialise NV · 510(k) 2026
- ConforMIS iTotal(R) Posterior Stabilized (PS) Knee Replacement System (iTotal PS)Conformis, Inc. · 510(k) 2016
- Zimmer Patient Specific Instruments/ System/PlannerMaterialise NV · 510(k) 2014
- Optetrak Total Knee SystemExactech, Inc. · 510(k) 2004
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.