510(k) device
AirFit F20
Resmed, Ltd. · Ventilator, Non-Continuous (Respirator) (BZD, Class II) · First FDA decision May 13, 2016, newest Jan 3, 2018Respiratory failure and ventilationSleep apnea
Last checked: Nov 19, 2026Source: FDA, CMS, ClinicalTrials.gov, PubMed
FDA decisions
2
510(k) pathway
Device trials
2
Naming "AirFit F20"
MAUDE reports
77
Brand "AirFit F20", since 2015
PubMed studies
0
Mentioning "AirFit F20"
AI summary across all sources
Written by AI from the FDA, CMS, ClinicalTrials.gov, MAUDE and PubMed data on this page, plus web searches for payment facts.
Product review deck
A slide deck reviewing AirFit F20
About ten slides on the technology, clinical evidence, regulatory history and commercialization, built with AI from the data on this page and web research, with sources. Download it as PowerPoint to edit or as PDF to share.
- Executive summary and key facts
- Technology
- Clinical evidence and trials
- Regulatory history and safety
- Reimbursement, market and competition
- Risks, outlook and sources
AI-written from public data and may contain mistakes. Check each fact against its source before relying on it. Not investment, legal, reimbursement or medical advice.
FDA decisions
The newest 2 of 2 decisions filed under this name.
- AirFit F20510(k)
K170924 · Decided Jan 3, 2018 · 280 days after receipt
Device type: Ventilator, Non-Continuous (Respirator) (BZD) · Anesthesiology & Respiratory
- AirFit F20510(k)
K153563 · Decided May 13, 2016 · 151 days after receipt
Device type: Ventilator, Non-Continuous (Respirator) (BZD) · Anesthesiology & Respiratory
Medicare coverage
National and local coverage documents whose titles match this device or its device type. Payment add-ons (NTAP, pass-through, TCET) and likely codes are covered in the summary above.
No Medicare coverage document has a title matching this device or device type. Many devices are paid under existing codes without one.
Clinical trials
Device studies on ClinicalTrials.gov that name "AirFit F20", any status and sponsor, newest update first.
- Trillium Validation StudyCompleted
Philips Clinical & Medical Affairs GlobalPublic (PHG)
NCT07444749 · 45 participants · Start Apr 21, 2026 · Primary completion May 28, 2026
Device: Trillium; AirFit F20 · For: Sleep Apnea, Obstructive
Sites: Fort Myers, Florida
Philips Clinical & Medical Affairs GlobalPublic (PHG)
NCT06932055 · 40 participants · Start Jul 15, 2025 · Primary completion Aug 19, 2025
Device: Trillium Full Face Mask; AirFit F20 · For: Obstructive Sleep Apnea
Sites: Lehigh Acres, Florida
Adverse event reports (MAUDE)
Reports received since 2015 naming the brand "AirFit F20", and for the whole product code for context.
This brand
77 reports: 39 malfunction, 32 injury, 6 death
Whole product code BZD
304,203 reports: 246,138 malfunction, 53,518 injury, 4,488 death
Newest reports for this brand
- AIRFIT F20 MED - AMERMalfunction
3004604967-2026-01407 · Received Oct 6, 2026 · Event dated Sep 8, 2026
Device type: Ventilator, Non-Continuous (Respirator) (BZD)
Device problem as coded: Insufficient Flow or Under Infusion
Patient problem as coded: Dyspnea
- AIRFIT F20 MED - AMERMalfunction
3007573469-2026-01407 · Received Oct 6, 2026 · Event dated Sep 8, 2026
Device type: Ventilator, Non-Continuous (Respirator) (BZD)
Device problem as coded: Insufficient Flow or Under Infusion
Patient problem as coded: Dyspnea
- AIRFIT F20 LGE - AMERMalfunction
3007573469-2026-01408 · Received Oct 6, 2026 · Event dated Sep 8, 2026
Device type: Ventilator, Non-Continuous (Respirator) (BZD)
Device problem as coded: Insufficient Flow or Under Infusion
Patient problem as coded: Dyspnea
2 more reports with a free account
Recalls
From FDA's recall database. FDA lists each recalled product separately, so one recall can appear several times. A recall is often a correction or a labeling change, not a removal from the market.
- AirFit F20 Full Face Mask and User GuideOpen, Classified
Z-0537-2024 · Started Nov 20, 2023 · Posted by FDA Dec 21, 2023
Reason given: Masks non-invasive interface for channeling airflow to patients have magnets that have potential magnetic interference with active medical implants that interact with magnets, and metallic implants/objects containing ferromagnetic material so contraindications and warnings about safe distance to medical...
Cause as FDA categorized it: Device Design · Device type: Ventilator, Non-Continuous (Respirator) (BZD)
- AirTouch F20 Full Face Mask and User GuideOpen, Classified
Z-0538-2024 · Started Nov 20, 2023 · Posted by FDA Dec 21, 2023
Reason given: Masks non-invasive interface for channeling airflow to patients have magnets that have potential magnetic interference with active medical implants that interact with magnets, and metallic implants/objects containing ferromagnetic material so contraindications and warnings about safe distance to medical...
Cause as FDA categorized it: Device Design · Device type: Ventilator, Non-Continuous (Respirator) (BZD)
- AirFit F30 Full Face Mask and User GuideOpen, Classified
Z-0541-2024 · Started Nov 20, 2023 · Posted by FDA Dec 21, 2023
Reason given: Masks non-invasive interface for channeling airflow to patients have magnets that have potential magnetic interference with active medical implants that interact with magnets, and metallic implants/objects containing ferromagnetic material so contraindications and warnings about safe distance to medical...
Cause as FDA categorized it: Device Design · Device type: Ventilator, Non-Continuous (Respirator) (BZD)
Published research
The most relevant PubMed records for "AirFit F20"[tiab]. The summary above synthesizes these.
No PubMed record mentions this device by name.
Other devices of this type
Also filed under product code BZD.
- K Series CPAP SystemsCurative Medical, Inc. · 510(k) 2014
- Mirage MicroResmed, Ltd. · 510(k) 2011
- AirFit N20Resmed, Ltd. · 510(k) 2017
- Mirage Full Face Mask Series 2Resmed, Ltd. · 510(k) 2002
- myAirResmed Corp · 510(k) 2025
- Mojo Full Face Vented MaskSleepnet Corporation · 510(k) 2024
- Mojo 2 Full Face Vented MaskSleepnet Corporation · 510(k) 2024
- iQ 2 Nasal Vented MaskSleepnet Corporation · 510(k) 2024
For information only. Nothing on this page is investment, legal, tax or medical advice, and nothing here is an offer to sell or a solicitation of an offer to buy any security. The data comes from public government sources, which can be incomplete, delayed or wrong. Check the original record before relying on it.