All In MedTech

Device safety

Medical device recalls: Class I, II and III explained, and how to look them up

Medical device recall lookup: what FDA Class I, II and III device recalls mean, how corrections and removals work under 21 CFR 806, and where to search.

By the All In MedTech editorial team · Published November 19, 2026 · 7 min read

A medical device recall is a correction or removal of a device that FDA considers to violate the law it administers. FDA sorts recalls into three classes by how much harm the device could cause. Class I is the most serious, Class II covers temporary or reversible harm, and Class III covers devices unlikely to cause any harm.

The class definitions come from FDA's general recall regulation at 21 CFR Part 7. Devices also have their own reporting rule, 21 CFR Part 806, on corrections and removals. This post explains both, then shows how to find recalls for a specific device or company.

The three recall classes

Section 7.3 of the regulation defines recall classification as the number, I, II or III, that FDA assigns to a recall "to indicate the relative degree of health hazard presented by the product being recalled." In plain terms:

ClassWhat it meansTypical examples
Class IA reasonable probability that using or being exposed to the device will cause serious adverse health consequences or death.A device that can fail without warning while it is keeping a patient alive.
Class IIUse may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.A component that can wear early, or labeling that could lead to a correctable error.
Class IIIUse is not likely to cause adverse health consequences.A labeling or packaging defect with no realistic path to harm.

Three points follow from the wording.

  • The class describes the hazard, not the size of the recall. A Class II recall can cover hundreds of thousands of units, and a Class I recall can cover a single lot.
  • The class is set by FDA, not by the company, after it reviews the health hazard. Until then a recall can show as not yet classified.
  • The class applies to the specific models, lots or serial numbers named in the recall. Other versions of the same device are not part of it.

Corrections and removals under Part 806

Device recalls differ from drug recalls in one practical way: many of them leave the device in place. Part 806 uses two terms.

  • Correction: the repair, modification, adjustment, relabeling, destruction or inspection (including patient monitoring) of a device without physically removing it from its point of use.
  • Removal: physically taking the device from its point of use to another place for repair, modification, adjustment, relabeling, destruction or inspection.

A software patch on an infusion pump, a new warning in the instructions for use, or a schedule of extra checks for patients with an implant are all corrections. Pulling unused stock from hospital shelves is a removal. FDA's What is a Medical Device Recall? page makes the same point: a recall does not always mean the device must stop being used or be returned.

What must be reported, and when

Under 21 CFR 806.10, a manufacturer or importer must send FDA a written report of any correction or removal it starts either to reduce a risk to health posed by the device, or to remedy a violation caused by the device that may present a risk to health. The report is due within 10 working days of starting the action. It must identify the device (including its unique device identifier where one applies), describe the problem and the actions taken, list any illnesses or injuries, and give the quantities and consignees involved.

"Risk to health" in Part 806 tracks the Class I and Class II definitions, which is why reportable corrections and removals tend to end up as Class I or II recalls.

Some actions are exempt from reporting under 806.1(b):

  • changes that improve performance or quality but do not reduce a risk to health or fix a violation;
  • market withdrawals, meaning a correction or removal involving a minor violation that would not be subject to FDA legal action, or no violation at all;
  • routine servicing, such as scheduled maintenance and replacing parts at the end of their normal life;
  • stock recoveries, where none of the affected units has left the manufacturer's control.

Even when no report is required, 806.20 requires the firm to keep a record of the correction or removal, including its justification for not reporting, for 2 years beyond the expected life of the device.

Who starts a recall

Most device recalls are started by the manufacturer. FDA can also request a recall under 21 CFR 7.45, and for devices it has a mandatory recall authority under section 518(e) of the Federal Food, Drug, and Cosmetic Act, set out in 21 CFR Part 810. FDA notes that it has rarely needed to order a device recall, because firms usually act first.

Each recall also has a status: open or ongoing while the firm carries out its strategy, and terminated once FDA decides that all reasonable efforts have been made to correct or remove the device.

Early alerts

In November 2024 FDA's device center announced a communications pilot to shorten the time between FDA learning of a potentially high-risk correction or removal and telling the public. It started with five clinical areas (cardiovascular, gastrorenal, general hospital, obstetrics and gynecology, and urology), and FDA updated the program in September 2025 to cover all medical devices.

Under the program FDA posts an early alert when it believes an action is likely to be among the most serious type of recall, before classification is finished. The Medical Device Recalls and Early Alerts page lists these alongside confirmed recalls. An early alert can later be confirmed as a recall, so treat it as a signal that information is still being gathered.

How to look up a device recall

FDA's Medical Device Recall Database

FDA's Medical Device Recall Database holds classified device recalls. You can search by product name, recalling firm, product code, recall class and date. Each record shows the reason for the recall, the affected models or lots, the distribution pattern, the action taken and the status.

The Enforcement Report

FDA's weekly Enforcement Report lists every recall FDA has classified across product types, including devices. It is a good place to see what was classified in a given week.

openFDA

The openFDA device recall API provides the same kind of records in machine-readable form. It is useful for counting recalls by firm, product code or year. openFDA states that its data are not validated for clinical decisions.

This site

Every device page on All In MedTech shows recall records next to the device's FDA decisions, Medicare coverage documents, trials and MAUDE report counts. Company pages gather recalls across all of a firm's devices.

Matching a recall to a device depends on the identifiers in the record, such as the product code and the 510(k) or PMA number. A few recalls lack these and will only be found by searching FDA's own databases.

Reading a recall record

These fields carry most of the information:

  1. Recall class: I, II or III, or not yet classified.
  2. Reason for recall: often the most useful field, for example a battery that can drain early or a software error that can display a wrong value.
  3. Product and identifiers: model numbers, lot or serial numbers and UDI device identifiers. Units outside these ranges are not included.
  4. Action: what the firm told customers to do, such as update software, inspect, stop using or return.
  5. Quantity and distribution: how many units and where they went.
  6. Dates and status: when the firm started the recall, when FDA posted it, and whether it is open or terminated.

A single recall event often covers several product records, one for each model or catalog number. Counts of records therefore overstate the number of separate recall events.

Recalls, adverse events and safety communications

A recall is one of several public safety signals for devices, and they answer different questions.

  • MAUDE reports describe individual adverse events and malfunctions reported to FDA. They can show a pattern before any recall, but a report alone does not prove the device caused the event. See what MAUDE is and how to search it.
  • Safety communications are FDA notices to patients and clinicians about a device risk. They may accompany a recall or stand alone.
  • Recalls describe a specific product problem and the firm's fix.

Looking at all three together, along with the device's FDA decision history, gives a fuller picture. That is the approach in our medtech due diligence checklist.

What a recall record does not tell you

Recall data describes the product problem and the corrective action. It does not say whether a specific patient was harmed or what a particular patient should do. The firm's notice and FDA's posting usually give instructions, and the treating clinician is the right person to advise on an implanted or prescribed device. Patients and clinicians can report a device problem to FDA through MedWatch.

New device recalls are classified every week, so check FDA's databases or a current device page rather than relying on a saved list.

Frequently asked questions

What is a Class I medical device recall?
It is the most serious recall class: FDA has found a reasonable probability that using or being exposed to the device will cause serious adverse health consequences or death.
Does a device recall mean I have to stop using the device?
Not always. Many device recalls are corrections, such as a software update, a labeling change or added patient monitoring, done without removing the device from where it is used.
How fast must a manufacturer tell FDA about a recall?
Under 21 CFR 806.10, a manufacturer or importer must report a correction or removal made to reduce a risk to health within 10 working days of starting it.
Where can I look up medical device recalls?
FDA's Medical Device Recall Database and weekly Enforcement Report list classified device recalls, and each device and company page on this site shows recall records from openFDA.

For information only, not medical, legal or investment advice. Facts about FDA, CMS and SEC rules are current as of the publication date; check the linked official source for changes.

Keep reading